Vibrations of the gradient coil support structure create sound waves. These are caused by the interactions of the magnetic field created by pulses of the current through the gradient coil with the main magnetic field in a manner similar to a loudspeaker coil. The sounds made by the scanner vary in volume and tone with the type of procedure being performed.
Sound pressure is reported on a logarithmic scale called sound-pressure level, expressed in decibel (dB) referenced to the weakest audible 1 000 Hz sound pressure of 2 * 10 -5 pascal (20 micropascal). Sound level meters contain filters that simulate the ear's frequency response. The most commonly used filter provides what is called 'A' weighting, with the letter 'A' appended to the dB units, i.e. dBA.
MRI system noise levels increase with field strength.
Disposable earplugs and/or headphones for the patient are recommended in high-field systems. Noise-canceling systems and special earphones are available, and active acoustic control systems were developed, e.g. softtone, pianissimo. A sequence with low noise gradient pulses is also called 'whisper sequence'.
See also Phon and Decibel. | | | | | • Share the entry 'Acoustic Noise': | | | | | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Cylindrical Wide Short Bore
Opt. (WIP) Single and Multi Voxel
SE, FE, IR, FastSE, FastIR, FastFLAIR, Fast STIR, FastFE, FASE, Hybrid EPI, Multi Shot EPI; Angiography: 2D(gate/non-gate)/3D TOF, SORS-STC
IMAGING MODES
Single, multislice, volume study
TE
8 msec min. SE; 1.2 msec min. FE
less than 0.015 (256x256)
1.0 min. 2-DFT: 0.2 min. 3-DFT
32-1024, phase;; 64-1024, freq.
65.5 cm, patient aperture
4050 kg (bare magnet incl. L-He)
COOLING SYSTEM TYPE
Closed-loop water-cooled
Liquid helium: approx. less than 0.05 L/hr
Passive, active, auto-active
| | | | • View the DATABASE results for 'Excelart AG™ with Pianissimo' (2).
| | | | |
| | | | | |
| |
|
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Cylindrical Wide Short Bore
SE, FE, IR, FastSE, FastIR, FastFLAIR, Fast STIR, FastFE, FASE, Hybrid EPI, Multi Shot EPI; Angiography: 2D(gate/non-gate)/3D TOF, SORS-STC
IMAGING MODES
Single, multislice, volume study
TE
8 msec min. SE; 0.9 msec min. FE
less than 0.011 (256x256)
1.0 min. 2-DFT: 0.2 min. 3-DFT
32-1024, phase;; 64-1024, freq.
65.5 cm, patient aperture
4050 kg (bare magnet incl. L-He)
POWER REQUIREMENTS
380/400/415/440/480 V
COOLING SYSTEM TYPE
Closed-loop water-cooled
Liquid helium: approx. less than 0.05 L/hr
Passive, active, auto-active
| | | | • View the DATABASE results for 'Excelart XG™ with Pianissimo' (2).
| | | | Further Reading: | News & More:
|
|
| |
| | | Searchterm 'Acoustic Noise' was also found in the following service: | | | | |
| | |
| |
|
(dB) A customary logarithmic measure most commonly used (in various ways) for measuring sound.
If one sound is 1 bel (10 decibel) 'louder' than another, this means the louder sound is 10 times louder than the fainter one. A difference of 20 decibel corresponds to an increase of 10 x 10 or 100 times in intensity.
For sound pressure (the pressure exerted by the sound waves) 0 decibel equals 20 micropascal (μPa), and for sound power 0 decibel sometimes equals 1 picowatt.
See also Phon and Acoustic Noise. | | | | • View the DATABASE results for 'Decibel' (4).
| | | | |
| | | | | |
| |
|
From Siemens Medical Systems;
MAGNETOM Avanto with Tim - Total imaging matrix technology.
For true whole-body anatomical coverage. For ultra-fast image
acquisition. Aids the physician in fast and precise
evaluation of systemic diseases like colon cancer, metastasis imaging, vessel diseases, and preventional exams. For claustrophobic patients,
MAGNETOM Avanto enables feetfirst exams for nearly all MR procedures. For obese patients, MAGNETOM Avanto supports up to 200 kg (400 lbs), without table movement restrictions. The AudioComfort technology enables up to a 30 dB(A) acoustic noise reduction, that means nearly all clinical routine sequences are running under 99 dB(A).
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Compact short bore
Body, Tim [32 x 8], Tim [76 x18], Tim [76 coil elements with up to 32 RF channels]
GRE, IR, FIR, STIR, TrueIR/FISP, FSE, FLAIR, MT, SS-FSE, MT-SE, MTC, MSE, EPI, 3D DESS//CISS/PSIF, GMR
IMAGING MODES
Single, multislice, volume study, multi angle, multi oblique
1024 x 1024 full screen display
POWER REQUIREMENTS
380/400/420/440/480 V
Passive, act.; 1st order std./2nd opt.
| | | | • View the DATABASE results for 'MAGNETOM Avanto™' (2).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | | |
| | 1 - 5 (of 10) nextResult Pages : [1] [2] |
| |
|
| |
| Look Ups |
| |