Contraindications | ![Forum -
related threads](gifs/t_ico_fo.gif) ![MRI Resource Directory:
- Safety -](gifs/dir_3.gif) |
| |
|
The principal contraindications of the MRI procedure are mostly related to the presence of metallic implants in a patient. The risks of MRI scans increase with the used field strength. In general, implants are becoming increasingly MR safe and an individual evaluation is carried out for each case.
Some patients should not be examined in MRI machines, or come closer than the 5 Gauss line to the system.
Absolute Contraindications for the MRI scan:
•
electronically, magnetically, and mechanically activated implants
•
metallic splinters in the eye
Patients with absolute contraindications should not be examined or only with special MRI safety precautions. Patients with an implanted cardiac pacemaker have been scanned on rare occasions, but pacemakers are generally considered an absolute contraindication. Relative contraindications may pose a relative hazard, and the type and location of an implant should be assessed prior to the MRI examination.
Relative Contraindications for the MRI scan:
•
other pacemakers, e.g. for the carotid sinus
Osteosynthesis material is usually anchored so well in the patients that no untoward effect will result. Another effect on metal parts in the patient's body is the heating of these parts through induction. In addition, image quality may be severely degraded. The presence of other metallic implants such as surgical clips etc. should be made known to the MRI operators. Most of these materials are non-magnetic, but if magnetic, they can pose a hazard.
See also MRI safety, Pregnancy, Claustrophobia and Tattoos. | | | | | | | | | • For this and other aspects of MRI safety see our InfoSheet about MRI Safety. | | | • Patient-related information is collected in our MRI Patient Information.
| | | | • Share the entry 'Contraindications': | | | | | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| ![](gifs/nb.gif) | ![](gifs/mores_2.gif) | Searchterm 'Contraindications' was also found in the following service: | | | | |
| ![](gifs/nb.gif) | ![](gifs/hg_1.gif) |
| |
|
| | ![spacer](gifs/nb.gif) | | • View the DATABASE results for 'Low Field MRI' (8).![Open this link in a new window](gifs/nw3_g.gif)
| | | • View the NEWS results for 'Low Field MRI' (5).![Open this link in a new window. Open this link in a new window.](gifs/nw3_g.gif)
| | | | Further Reading: | Basics:
|
|
News & More:
| ![](gifs/triangle_blue_1.gif) |
Safety of Bedside Portable Low-Field Brain MRI in ECMO Patients Supported on Intra-Aortic Balloon Pump Friday, 18 November 2022 by www.mdpi.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Researchers at the University of Tsukuba develop a portable MRI system specifically for identifying wrist cartilage damage among athletes, providing a convenient means of early detection and treatment of injuries Tuesday, 26 April 2022 by www.tsukuba.ac.jp | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
This bizarre looking helmet can create better brain scans Friday, 11 February 2022 by www.sciencedaily.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
A low-cost and shielding-free ultra-low-field brain MRI scanner Tuesday, 14 December 2021 by www.nature.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Portable MRI provides life-saving information to doctors treating strokes Thursday, 5 August 2021 by news.yale.edu | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Synaptive Evry, an MRI for Any Space, Cleared by FDA Thursday, 30 April 2020 by www.medgadget.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
World's First Portable MRI Cleared by FDA Monday, 17 February 2020 by www.medgadget.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Introducing a point-of-care MRI system Tuesday, 29 October 2019 by healthcare-in-europe.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Opportunities in Interventional and Diagnostic Imaging by Using High-performance Low-Field-Strength MRI Tuesday, 1 October 2019 by pubs.rsna.org | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Portable 'battlefield MRI' comes out of the lab Thursday, 30 April 2015 by physicsworld.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Portable MRI could aid wounded soldiers and children in the third world Thursday, 23 April 2015 by phys.org |
|
| |
| ![](gifs/nb.gif) | ![](gifs/link_ldir_001.gif) | MRI Safety Resources | | | | |
| ![](gifs/nb.gif) | ![](gifs/nb.gif) | ![](gifs/hg_1.gif) |
| |
|
It is important to remember when working around a superconducting magnet that the magnetic field is always on. Under usual working conditions the field is never turned off. Attention must be paid to keep all ferromagnetic items at an adequate distance from the magnet. Ferromagnetic objects which came accidentally under the influence of these strong magnets can injure or kill individuals in or nearby the magnet, or can seriously damage every hardware, the magnet itself, the cooling system, etc..
See MRI resources Accidents.
The doors leading to a magnet room should be closed at all times except when entering or exiting the room. Every person working in or entering the magnet room or adjacent rooms with a magnetic field has to be instructed about the dangers. This should include the patient, intensive-care staff, and maintenance-, service- and cleaning personnel, etc..
The 5 Gauss limit defines the 'safe' level of static magnetic field exposure. The value of the absorbed dose is fixed by the authorities to avoid heating of the patient's tissue and is defined by the specific absorption rate.
Leads or wires that are used in the magnet bore during imaging procedures, should not form large-radius wire loops. Leg-to-leg and leg-to-arm skin contact should be prevented in order to avoid the risk of burning due to the generation of high current loops if the legs or arms are allowed to touch. The patient's skin should not be in contact with the inner bore of the magnet.
The outflow from cryogens like liquid helium is improbable during normal operation and not a real danger for patients.
The safety of MRI contrast agents is tested in drug trials and they have a high compatibility with very few side effects. The variations of the side effects and possible contraindications are similar to X-ray contrast medium, but very rare. In general, an adverse reaction increases with the quantity of the MRI contrast medium and also with the osmolarity of the compound.
See also 5 Gauss Fringe Field, 5 Gauss Line, Cardiac Risks, Cardiac Stent, dB/dt, Legal Requirements, Low Field MRI, Magnetohydrodynamic Effect, MR Compatibility, MR Guided Interventions, Claustrophobia, MRI Risks and Shielding. | | | | | | | | | • View the DATABASE results for 'MRI Safety' (42).![Open this link in a new window](gifs/nw3_g.gif)
| | | • View the NEWS results for 'MRI Safety' (13).![Open this link in a new window. Open this link in a new window.](gifs/nw3_g.gif)
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| ![](gifs/nb.gif) | ![](gifs/mores_2.gif) | Searchterm 'Contraindications' was also found in the following services: | | | | |
| ![](gifs/nb.gif) | ![](gifs/hg_1.gif) |
| |
|
In the last years, cardiac MRI techniques have progressively improved. No other noninvasive imaging modality provides the same degree of contrast and temporal resolution for the assessment of cardiovascular anatomy and pathology. Contraindications MRI are the same as for other magnetic resonance techniques.
The primary advantage of MRI is extremely high contrast resolution between different tissue types, including blood. Moreover, MRI is a true 3 dimensional imaging modality and images can be obtained in any oblique plane along the true cardiac axes while preserving high temporal and spatial resolution with precise demonstration of cardiac anatomy without the administration of contrast media.
Due to these properties, MRI can precisely characterize cardiac function and quantify cavity volumes, ejection fraction, and left ventricular mass. In addition, cardiac MRI has the ability to quantify flow (see flow quantification), including bulk flow in vessels, pressure gradients across stenosis, regurgitant fractions and shunt fractions. Valve morphology and area can be determined and the severity of stenosis quantified. In certain disease states, such as myocardial infarction, the contrast resolution of MRI is further improved by the addition of extrinsic contrast agents (see myocardial late enhancement).
A dedicated cardiac coil, and a field strength higher than 1 Tesla is recommended to have sufficient signal. Cardiac MRI acquires ECG gating. Cardiac gating (ECGs) obtained within the MRI scanner, can be degraded by the superimposed electrical potential of flowing blood in the magnetic field. Therefore, excellent contact between the skin and ECG leads is necessary. For male patients, the skin at the lead sites can be shaved. A good cooperation of the patient is necessary because breath holding at the end of expiration is practiced during the most sequences.
See also Displacement Encoding with Stimulated Echoes.
For Ultrasound Imaging (USI) see Cardiac Ultrasound at Medical-Ultrasound-Imaging.com.
See also the related poll results: ' In 2010 your scanner will probably work with a field strength of' and ' MRI will have replaced 50% of x-ray exams by' | | | | ![spacer](gifs/nb.gif) | | • View the DATABASE results for 'Cardiac MRI' (15).![Open this link in a new window](gifs/nw3_g.gif)
| | | • View the NEWS results for 'Cardiac MRI' (15).![Open this link in a new window. Open this link in a new window.](gifs/nw3_g.gif)
| | | | Further Reading: | Basics:
|
|
News & More:
| ![](gifs/triangle_blue_1.gif) |
MRI technology visualizes heart metabolism in real time Friday, 18 November 2022 by medicalxpress.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Even early forms of liver disease affect heart health, Cedars-Sinai study finds Thursday, 8 December 2022 by www.eurekalert.org | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
MRI sheds light on COVID vaccine-associated heart muscle injury Tuesday, 15 February 2022 by www.sciencedaily.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Radiologists must master cardiac CT, MRI to keep pace with demand: The heart is not a magical organ Monday, 1 March 2021 by www.radiologybusiness.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Diffusion weighted imaging (DWI) and diffusion tensor imaging (DTI) in the heart (myocardium) Sunday, 30 August 2020 by github.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Non-invasive diagnostic procedures for suspected CHD: Search reveals informative evidence Wednesday, 8 July 2020 by medicalxpress.co | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Cardiac MRI Becoming More Widely Available Thanks to AI and Reduced Exam Times Wednesday, 19 February 2020 by www.dicardiology.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Controlling patient's breathing makes cardiac MRI more accurate Friday, 13 May 2016 by www.upi.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
Precise visualization of myocardial injury: World's first patient-based cardiac MRI study using 7T MRI Wednesday, 10 February 2016 by medicalxpress.com | ![](gifs/nb.gif) | ![](gifs/triangle_blue_1.gif) |
New technique could allow for safer, more accurate heart scans Thursday, 10 December 2015 by www.gizmag.com |
|
| |
| ![](gifs/nb.gif) | ![](gifs/mores_2.gif) | Searchterm 'Contraindications' was also found in the following service: | | | | |
| ![](gifs/nb.gif) | ![](gifs/hg_1.gif) |
| |
|
During the MRI scan an augmentation of T waves is observed at fields used in standard imaging but this possible MRI side effect is completely reversible upon removal from the magnet. A field strength dependent increase in the amplitude of the ECG in rats has been observed during exposure to high homogeneous stationary magnetic fields, but this side effect is not transferable to standard imaging situations for humans.
The minimum level at which augmentation can be observed is 0.3 T and increases by higher field strength.
An augmentation in T-wave amplitude can occur instantaneously and is immediately reversible after exposure to the magnetic field ceased. There should be no abnormalities in the ECG in the later follow-up. Augmentation of the signal amplitude in the T-wave segment may result from superimposed electrical potential.
No circulatory alterations coincide with the ECG changes. Therefore, no biological risks are believed to be associated with them.
For more MRI safety information see also Contraindications
and MRI Risks. | | ![spacer](gifs/nb.gif) | | • View the DATABASE results for 'Cardiac Risks' (2).![Open this link in a new window](gifs/nw3_g.gif)
| | | | Further Reading: | Basics:
|
|
| |
| ![](gifs/nb.gif) | | ![](gifs/nb.gif) | ![](gifs/nb.gif) |
| ![](gifs/nb.gif) | 1 - 5 (of 9) nextResult Pages : [1] [2] |
| |
|
| |
![](gifs/bs_2.gif) | Look Ups |
| |