Saturation Recovery
 
(SR) Particular type of partial saturation pulse sequence in which the preceding pulses leave the spins in a state of saturation, so that recovery at the time of the next pulse has taken place from an initial condition of no magnetization. A rare used MRI pulse sequence that generates a predominantly proton density dependent signal, basically employing a 90° RF excitation pulse, with a very long repetition time. With this technique T1 times can be measured faster than with inversion recovery pulse sequences.
This saturation recovery sequence consists of multiple 90° radio frequency (RF) pulses with a short repetition time. A spoiler gradient pulse dephases the longitudinal magnetization that remains after the first 90° radio frequency pulse. A repetition time interval after the application of this spoiling gradient turns an additional 90° pulse the new developed longitudinal magnetization into the transverse plane, followed by recording a gradient echo.
spacer
 
• Related Searches:
    • Gradient Echo Sequence
    • Suppression
    • Progressive Saturation
    • Partial Saturation
    • Chemical Shift Imaging
 
Further Reading:
  Basics:
Contrast mechanisms in magnetic resonance imaging
2004   by www.iop.org    
Searchterm 'Excitation' was also found in the following services: 
spacer
Radiology  (5) Open this link in a new windowUltrasound  (6) Open this link in a new window
Scan TimeForum -
related threads
 
(SCT) The total scan time is the time required to collect all data needed to generate the programmed images. The scan time is related to the used pulse sequence and dependent on the assemble of parameters like e.g., repetition time (TR), Matrix, number of signal averages (NSA), TSE- or EPI factor and flip angle.
For example, the total scan time for a standard spin echo or gradient echo sequence is number of repetitions x the scan time per repetition (means the product of repetition time (TR), number of phase encoding steps, and NSA).

See also Number of Excitations, Turbo Spin Echo Turbo Factor, Echo Planar Imaging Factor, Flip Angle and Image Acquisition Time.

See also acronyms for 'scan time parameters' from different manufacturers.
spacer

• View the DATABASE results for 'Scan Time' (47).Open this link in a new window


• View the NEWS results for 'Scan Time' (10).Open this link in a new window.
 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
  News & More:
For MRI, time is of the essence A new generation of contrast agents could make for faster and more accurate imaging
Tuesday, 28 June 2011   by scienceline.org    
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Bioinformatics - Shoulder MRI - Stimulator pool - Implant and Prosthesis - MR Guided Interventions - Absorption and Emission
 
Selective Irradiation
 
spacer
Searchterm 'Excitation' was also found in the following services: 
spacer
News  (1)  Forum  (11)  
 
Sensitive Point
 
The sensitive point in an object (or a region) is the point from which information is being measured, as in sequential point imaging or in single voxel MR spectroscopy sequential point imaging used. To define a sensitive point, it must be selected in three dimensions. Different combinations of selective excitation, inversion, or saturation pulses can be used. Sometimes, combinations of two or more measurements are necessary to define the sensitive point (or volume).

See also Magnetic Resonance Spectroscopy.
spacer

• View the DATABASE results for 'Sensitive Point' (3).Open this link in a new window

Searchterm 'Excitation' was also found in the following services: 
spacer
Radiology  (5) Open this link in a new windowUltrasound  (6) Open this link in a new window
Sequential Line Imaging
 
MR imaging techniques in which the image is built up from successive lines through the object. In various schemes, the lines are isolated by oscillating magnetic field gradients or selective excitation, and then the NMR signals from the selected line are encoded for position by detecting the FID or spin echo in the presence of a magnetic field gradient along the line; the Fourier transformation of the detected signal then yields the distribution of emitted NMR signal along the line.
spacer

• View the DATABASE results for 'Sequential Line Imaging' (4).Open this link in a new window

MRI Resources 
Crystallography - Contrast Enhanced MRI - Equipment - Stent - Guidance - MRI Training Courses
 
previous      76 - 80 (of 96)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]