Phase Wrapping ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Parts of the body, which extend beyond the Field of view in the phase encoding direction may still be visible in the image as artifacts.
Please note that aliasing in the frequency direction is also possible but can be eliminated in the same manner or by applying frequency specific filters to the received signal.

See Aliasing Artifact.
spacer
 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
Searchterm 'Field Of View' was also found in the following services: 
spacer
Radiology  (10) Open this link in a new windowUltrasound  (4) Open this link in a new window
Phased Array CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
The phased array coils operate typically as receive only coils. In that case, the in the MRI device implemented body coil act as the transmitter and sends the radio frequency energy to generate the excitation pulses. State-of-the-art array coil systems include the use of 4 (up to 32) coils with separate receivers. This method is often referred to as a phased array system, although the signals are not added such that the signal phase information is included. The use of phased array coils allows the decreasing of the number of signal averages, which shortens the scan time by high SNR and resolution.
High-sensitivity RF surface coils and digital processing algorithms have been developed that speed up image acquisition and reconstruction during the MRI scan.
Fast parallel imaging techniques, for example sensitivity encoding (SENSE), 'Partially Parallel Imaging with Localized Sensitivity' (PILS), Simultaneous Acquisition of Spatial Harmonics (SMASH) or Array Spatial Sensitivity Encoding Technique (ASSET) use phased array multichannel coils to further improve spatial and temporal resolution. The sensitivity profile of a phased array coil element is measured by a separate low resolution 3D acquisition over the entire field of view in the case of a SENSE acquisition. For an mSENSE measurement, a self-calibration acquires some of the missing lines in the center of the k-space.
Also called linear array coil or synergy surface coil.

See also the related poll result: '3rd party coils are better than the original manufacturer coils'
spacer

• View the DATABASE results for 'Phased Array Coil' (9).Open this link in a new window

 
Further Reading:
  Basics:
Coil Arrays for Parallel MRI: Introduction and Overview.
   by www.mr.ethz.ch    
MRI Resources 
Breast Implant - Equipment - Implant and Prosthesis - Resources - RIS - Pregnancy
 
Quadrature Detector
 
The quadrature detector is a part of the receiver that converts the high-frequency MRI signal to a lower frequency. This phase sensitive detector or demodulator detects the components of the signal in phase with a reference signal and 90° out of phase with the reference signal. The modulated signal contains i.e. the frequency range across the field of view encoded by the frequency encoding gradient. This may be performed by either analog or digital means.
spacer

• View the DATABASE results for 'Quadrature Detector' (7).Open this link in a new window

Searchterm 'Field Of View' was also found in the following services: 
spacer
News  (6)  Resources  (2)  Forum  (3)  
 
Radio Frequency Quadrature ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Radio frequency quadrature
DESCRIPTION
Bright spot in the center of the image, ghosting
REASON
RF detection circuitry problem
HELP
Call the service
Radio frequency quadrature artifacts occur when the detector channels of the quadrature detector are disturbed.
A DC offset of one amplifier output can e.g., produce a bright point in the center of the field of view (see also central point artifact), or a higher gain of one detector channel can generate diagonally ghosting.
mri safety guidance
Image Guidance
Radio frequency quadrature artifacts are technical faults and must be eliminated by the service.
spacer
 
Further Reading:
  News & More:
Magnetic Resonance Imaging (MRI)
2003   by www.hull.ac.uk    
Searchterm 'Field Of View' was also found in the following services: 
spacer
Radiology  (10) Open this link in a new windowUltrasound  (4) Open this link in a new window
ResolutionForum -
related threads
 
Resolution is a function of slice thickness, field of view (FOV) and matrix size. The resolution in plane is a function of FOV / matrix size.
spacer

• View the DATABASE results for 'Resolution' (150).Open this link in a new window


• View the NEWS results for 'Resolution' (10).Open this link in a new window.
 
Further Reading:
  Basics:
Micro-MRI Principles, Strengths, and Weaknesses
Tuesday, 10 January 2017   by www.news-medical.net    
  News & More:
High-resolution MRI enables direct imaging of neuronal activity - DIANA – direct imaging of neuronal activity
Friday, 18 November 2022   by physicsworld.com    
Deep learning-based single image super-resolution for low-field MR brain images
Saturday, 16 April 2022   by www.nature.com    
A Comparison of Methods for High-Spatial-Resolution Diffusion-weighted Imaging in Breast MRI
Tuesday, 25 August 2020   by pubs.rsna.org    
MRI Resources 
Lung Imaging - MRI Physics - Crystallography - Contrast Enhanced MRI - Brain MRI - Collections
 
previous      21 - 25 (of 29)     next
Result Pages : [1]  [2 3 4 5 6]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]