Field StrengthForum -
related threads
 
In every MR examination, a large static magnetic field is applied. Field strengths for clinical equipment can vary between 0.2 and 3 T; experimental imaging units have a field strength of up to 11 T, depending on the MRI equipment used. In MRS, field strengths up to 12 T are currently used. The field strength of the magnet will influence the quality of the MR image regarding chemical shift artifacts, the signal to noise ratio (SNR), motion sensitivity and susceptibility artifacts.

See also the related poll result: 'In 2010 your scanner will probably work with a field strength of'
spacer
 
• Share the entry 'Field Strength':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • 5 Gauss Line
    • Susceptibility Artifact
    • Bandwidth
    • Low Field MRI
    • B1
 
Further Reading:
  Basics:
What affects the strength of a magnet?
   by my.execpc.com    
Magnetic Field of the Strongest Magnet
2003   by hypertextbook.com    
  News & More:
Impact of Magnetic Field Inhomogeneity on the Quality of Magnetic Resonance Images and Compensation Techniques: A Review
Saturday, 1 October 2022   by www.dovepress.com    
Low Power MRI Helps Image Lungs, Brings Costs Down
Thursday, 10 October 2019   by www.medgadget.com    
Opportunities in Interventional and Diagnostic Imaging by Using High-performance Low-Field-Strength MRI
Tuesday, 1 October 2019   by pubs.rsna.org    
Low-field MRI: An MR physics perspective
Sunday, 13 January 2019   by onlinelibrary.wiley.com    
Ultra-high-field MRI allows for earlier diagnosis of multiple sclerosis
Friday, 4 May 2007   by www.eurekalert.org    
Low-intensity MRI takes first scan of a human brain
Wednesday, 14 November 2007   by technology.newscientist.com    
Searchterm 'Field Strength' was also found in the following services: 
spacer
Radiology  (5) Open this link in a new windowUltrasound  (2) Open this link in a new window
B0
 
[B0] A conventional symbol for the main magnetic field strength (magnetic flux density or induction) in a MRI system. Although historically used, H0 (units of magnetic field strength, ampere//meter) should be distinguished from the more appropriate B0 [units of magnetic induction, tesla].
In current MR systems it has a constant value over time varying from 0.02 to 4 T. Field strengths of 0.5 T and above are generated with superconductive magnets. High field strengths have a better signal to noise ratio (SNR). The optimal imaging field strength for clinical imaging is between 0.5 and 2.0 T.

See also the related poll result: 'In 2010 your scanner will probably work with a field strength of'
spacer

• View the DATABASE results for 'B0' (41).Open this link in a new window

 
Further Reading:
  Basics:
Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities.
Tuesday, 1 August 2006   by www.ncbi.nlm.nih.gov    
Magnetic Field
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
Turbo-FLASH Based Arterial Spin Labeled Perfusion MRI at 7 T
Thursday, 20 June 2013   by www.plosone.org    
Penn researchers to get 7 Tesla whole-body MRI system
Monday, 28 August 2006   by www.eurekalert.org    
Optimizing SPIR and SPAIR fat suppression
Tuesday, 30 November 2004   by clinical.netforum.healthcare.philips.com    
MRI Resources 
Veterinary MRI - Musculoskeletal and Joint MRI - Mobile MRI Rental - Calculation - Homepages - Open Directory Project
 
MRI RisksMRI Resource Directory:
 - Safety -
 
The subacute risks and side effects of magnetic and RF fields (for patients and staff) have been intensively examined for a long time, but there have been no long-term studies following persons who have been exposed to the static magnetic fields used in MRI. However, no permanent hazardous effects of a static magnetic field exposure upon human beings have yet been demonstrated.
Temporary possible side effects of high magnetic and RF fields:
Varying magnetic fields can induce so-called magnetic phosphenes that occur when an individual is subject to rapid changes of 2-5 T/s, which can produce a flashing sensation in the eyes. This temporary side effect does not seem to damage the eyes. Static field strengths used for clinical MRI examinations vary between 0.2 and 3.0 tesla;; field changes during the MRI scan vary in the dimension of mT/s. Experimental imaging units can use higher field strengths of up to 14.0 T, which are not approved for human use.
The Radio frequency pulses mainly produce heat, which is absorbed by the body tissue. If the power of the RF radiation is very high, the patient may be heated too much. To avoid this heating, the limit of RF exposure in MRI is up to the maximum specific absorption rate (SAR) of 4 W/kg whole body weight (can be different from country to country). For MRI safety reasons, the MRI machine starts no sequence, if the SAR limit is exceeded.
Very high static magnetic fields are needed to reduce the conductivity of nerves perceptibly. Augmentation of T waves is observed at fields used in standard imaging but this side effect in MRI is completely reversible upon removal from the magnet. Cardiac arrhythmia threshold is typically set to 7-10 tesla. The magnetohydrodynamic effect, which results from a voltage occurring across a vessel in a magnetic field and percolated by a saline solution such as blood, is irrelevant at the field strengths used.

The results of some animal and cellular studies suggest the possibility that electromagnetic fields may act as co-carcinogens or tumor promoters, but the data are inconclusive. Up to 45 tesla, no important effects on enzyme systems have been observed. Neither changes in enzyme kinetics, nor orientation changes in macromolecules have been conclusively demonstrated.
There are some publications associating an increase in the incidence of leukemia with the location of buildings close to high-current power lines with extremely low-frequency (ELF) electromagnetic radiation of 50-60 Hz, and industrial exposure to electric and magnetic fields but a transposition of such effects to MRI or MRS seems unlikely.
Under consideration of the MRI safety guidelines, real dangers or risks of an exposure with common MRI field strengths up to 3 tesla as well as the RF exposure during the MRI scan, are not to be expected.

For more MRI safety information see also Nerve Conductivity, Contraindications, Pregnancy and Specific Absorption Rate.

See also the related poll result: 'In 2010 your scanner will probably work with a field strength of'
spacer

• View the DATABASE results for 'MRI Risks' (9).Open this link in a new window


• View the NEWS results for 'MRI Risks' (3).Open this link in a new window.
 
Further Reading:
  Basics:
MRI in Patients with Implanted Devices: Current Controversies
Monday, 1 August 2016   by www.acc.org    
Working with MRI machines may cause vertigo: Study
Wednesday, 25 June 2014   by www.cos-mag.com    
Physics of MRI Safety
   by www.aapm.org    
When Your Kid Needs an MRI: Optimizing the Experience
Tuesday, 29 March 2016   by health.usnews.com    
  News & More:
How safe is 7T MRI for patients with neurosurgical implants?
Thursday, 17 November 2022   by healthimaging.com    
CT contrast reaction raises MRI contrast risk
Tuesday, 22 February 2022   by www.sciencedaily.com    
CSU study explores MRI distress and patient experience
Thursday, 7 May 2020   by www.portnews.com.au    
Noise from Magnetic Resonance Imaging Can Have Short-Term Impact on Hearing
Thursday, 22 February 2018   by www.diagnosticimaging.com    
Women with permanent make-up tattoos suffer horrific facial burns after going in for MRI scans - which create an electric current in the ink
Monday, 4 July 2016   by www.dailymail.co.uk    
FDA Dials in on MRI Safety of Passive Implantable Medical Devices
Wednesday, 24 June 2015   by www.raps.org    
Searchterm 'Field Strength' was also found in the following services: 
spacer
News  (13)  Resources  (3)  Forum  (7)  
 
DeviceForum -
related threadsInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.
 
Magnetic resonance imaging (MRI) is based on the magnetic resonance phenomenon, and is used for medical diagnostic imaging since ca. 1977 (see also MRI History).
The first developed MRI devices were constructed as long narrow tunnels. In the meantime the magnets became shorter and wider. In addition to this short bore magnet design, open MRI machines were created. MRI machines with open design have commonly either horizontal or vertical opposite installed magnets and obtain more space and air around the patient during the MRI test.
The basic hardware components of all MRI systems are the magnet, producing a stable and very intense magnetic field, the gradient coils, creating a variable field and radio frequency (RF) coils which are used to transmit energy and to encode spatial positioning. A computer controls the MRI scanning operation and processes the information.
The range of used field strengths for medical imaging is from 0.15 to 3 T. The open MRI magnets have usually field strength in the range 0.2 Tesla to 0.35 Tesla. The higher field MRI devices are commonly solenoid with short bore superconducting magnets, which provide homogeneous fields of high stability.
There are this different types of magnets:
The majority of superconductive magnets are based on niobium-titanium (NbTi) alloys, which are very reliable and require extremely uniform fields and extreme stability over time, but require a liquid helium cryogenic system to keep the conductors at approximately 4.2 Kelvin (-268.8° Celsius). To maintain this temperature the magnet is enclosed and cooled by a cryogen containing liquid helium (sometimes also nitrogen).
The gradient coils are required to produce a linear variation in field along one direction, and to have high efficiency, low inductance and low resistance, in order to minimize the current requirements and heat deposition. A Maxwell coil usually produces linear variation in field along the z-axis; in the other two axes it is best done using a saddle coil, such as the Golay coil.
The radio frequency coils used to excite the nuclei fall into two main categories; surface coils and volume coils. The essential element for spatial encoding, the gradient coil sub-system of the MRI scanner is responsible for the encoding of specialized contrast such as flow information, diffusion information, and modulation of magnetization for spatial tagging.
An analog to digital converter turns the nuclear magnetic resonance signal to a digital signal. The digital signal is then sent to an image processor for Fourier transformation and the image of the MRI scan is displayed on a monitor.

For Ultrasound Imaging (USI) see Ultrasound Machine at Medical-Ultrasound-Imaging.com.

See also the related poll results: 'In 2010 your scanner will probably work with a field strength of' and 'Most outages of your scanning system are caused by failure of'
Radiology-tip.comradGamma Camera,  Linear Accelerator
spacer
Medical-Ultrasound-Imaging.comUltrasound Machine,  Real-Time Scanner
spacer

• View the DATABASE results for 'Device' (141).Open this link in a new window


• View the NEWS results for 'Device' (29).Open this link in a new window.
 
Further Reading:
  News & More:
small-steps-can-yield-big-energy-savings-and-cut-emissions-mris
Thursday, 27 April 2023   by www.itnonline.com    
Portable MRI can detect brain abnormalities at bedside
Tuesday, 8 September 2020   by news.yale.edu    
Point-of-Care MRI Secures FDA 510(k) Clearance
Thursday, 30 April 2020   by www.diagnosticimaging.com    
World's First Portable MRI Cleared by FDA
Monday, 17 February 2020   by www.medgadget.com    
Low Power MRI Helps Image Lungs, Brings Costs Down
Thursday, 10 October 2019   by www.medgadget.com    
Cheap, portable scanners could transform brain imaging. But how will scientists deliver the data?
Tuesday, 16 April 2019   by www.sciencemag.org    
The world's strongest MRI machines are pushing human imaging to new limits
Wednesday, 31 October 2018   by www.nature.com    
Kyoto University and Canon reduce cost of MRI scanner to one tenth
Monday, 11 January 2016   by www.electronicsweekly.com    
A transportable MRI machine to speed up the diagnosis and treatment of stroke patients
Wednesday, 22 April 2015   by medicalxpress.com    
Portable 'battlefield MRI' comes out of the lab
Thursday, 30 April 2015   by physicsworld.com    
Chemists develop MRI technique for peeking inside battery-like devices
Friday, 1 August 2014   by www.eurekalert.org    
New devices doubles down to detect and map brain signals
Monday, 23 July 2012   by scienceblog.com    
Searchterm 'Field Strength' was also found in the following services: 
spacer
Radiology  (5) Open this link in a new windowUltrasound  (2) Open this link in a new window
Knee MRI
 
Knee MRI, with its high soft tissue contrast is one of the main imaging tools to depict knee joint pathology. MRI allows accurate imaging of intra-articular structures such as ligaments, cartilage, menisci, bone marrow, synovium, and adjacent soft tissue.
Knee exams require a dedicated extremity coil, providing a homogenous imaging volume and high SNR to ensure best signal coverage. A complete knee MR examination includes for example sagittal and coronal T1 weighted, and proton density weighted pulse sequences +/- fat saturation, or STIR sequences. For high spatial resolution, maximal 4 mm thick slices with at least an in plane resolution of 0.75 mm and small gap are recommended. To depict the anterior cruciate ligament clearly, the sagittal plane has to be rotated 10 - 20° externally (parallel to the medial border of the femoral condyle). Retropatellar cartilage can bee seen for example in axial T2 weighted gradient echo sequences with Fatsat. However, the choice of the pulse sequences is depended of the diagnostic question, the used scanner, and preference of the operator.
Diagnostic quality in knee imaging is possible with field strengths ranging from 0.2 to 3T. With low field strengths more signal averages must be measured, resulting in increased scan times to provide equivalent quality as high field strengths.
More diagnostic information of meniscal tears and chondral defects can be obtained by direct magnetic resonance arthrography, which is done by introducing a dilute solution of gadolinium in saline (1:1000) into the joint capsule. The knee is then scanned in all three planes using T1W sequences with fat suppression. For indirect arthrography, the contrast is given i.v. and similar scans are started 20 min. after injection and exercise of the knee.
Frequent indications of MRI scans in musculoskeletal knee diseases are:
e.g., meniscal degeneration and tears, ligament injuries, osteochondral fractures, osteochondritis dissecans, avascular bone necrosis and rheumatoid arthritis.

See also Imaging of the Extremities and STIR.
 
Images, Movies, Sliders:
 Sagittal Knee MRI Images T1 Weighted  Open this link in a new window
      

 Anatomic MRI of the Knee 2  Open this link in a new window
    
SlidersSliders Overview

 Knee MRI Coronal Pd Spir 001  Open this link in a new window
 Sagittal Knee MRI Images STIR  Open this link in a new window
      

 Axial Knee MRI Images T2 Weighted  Open this link in a new window
 Anatomic MRI of the Knee 1  Open this link in a new window
    
SlidersSliders Overview

 
spacer

• View the DATABASE results for 'Knee MRI' (4).Open this link in a new window


• View the NEWS results for 'Knee MRI' (4).Open this link in a new window.
 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
Knee, Anterior Cruciate Ligament Injuries (MRI)
Tuesday, 28 March 2006   by www.emedicine.com    
  News & More:
NSAIDs May Worsen Arthritis Inflammation
Monday, 21 November 2022   by www.itnonline.com    
A Knee MRI in Half the Time? It's Possible
Thursday, 8 April 2021   by www.diagnosticimaging.com    
Seniors, patients, astronauts will all benefit from new USask research on bone health
Saturday, 27 February 2021   by www.yorktonthisweek.com    
3D mapping algorithm reads knee MRIs for new arthritis treatments
Thursday, 11 June 2020   by www.healthimaging.com    
MRI T2 Mapping of the Knee Providing Synthetic Morphologic Images: Comparison to Conventional Turbo Spin-Echo MRI
Tuesday, 1 October 2019   by pubs.rsna.org    
Researcher uses MRI to measure joint's geometry and role in severe knee injury
Tuesday, 23 September 2014   by medicalxpress.com    
Abnormalities on MRI predict knee replacement
Monday, 9 March 2015   by medicalxpress.com    
MRI Resources 
Contrast Agents - Spectroscopy - IR - Jobs pool - General - Examinations
 
     1 - 5 (of 136)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]