Field of View
 
(FOV) Defined as the size of the two or three dimensional spatial encoding area of the image. Usually defined in units of mm². The FOV is the square image area that contains the object of interest to be measured. The smaller the FOV, the higher the resolution and the smaller the voxel size but the lower the measured signal. Useful for decreasing the scantime is a field of view different in the frequency and phase encoding directions (rectangular field of view - RFOV).
The magnetic field homogeneity decreases as more tissue is imaged (greater FOV). As a result the precessional frequencies change across the imaging volume. That can be a problem for fat suppression imaging. This fat is precessing at the expected frequency only in the center of the imaging volume. E.g. frequency specific fat saturation pulses become less effective when the field of view is increased. It is best to use smaller field of views when applying fat saturation pulses.
mri safety guidance
Image Guidance
Smaller FOV required higher gradient strength and concludes low signal. Therefore you have to find a compromise between these factors. The right choice of the field of view is important for MR image quality. When utilizing small field of views and scanning at a distance from the isocenter (more problems with artifacts) it is obviously important to ensure that the region of interest is within the scanning volume.
A smaller FOV in one direction is available with the function rectangular field of view (RFOV).

See also Field Inhomogeneity Artifact.
 
Images, Movies, Sliders:
 MRI - Anatomic Imaging of the Foot  Open this link in a new window
    
SlidersSliders Overview

 MRI - Anatomic Imaging of the Ankle 1  Open this link in a new window
    
SlidersSliders Overview

 
spacer
 
• Share the entry 'Field of View':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Precession
    • Phase Encoding
    • Homogeneity
    • Device
    • Signal to Noise Ratio
 
Further Reading:
  Basics:
Image Characteristics and Quality
   by www.sprawls.org    
  News & More:
Optimizing Musculoskeletal MR
   by rad.usuhs.mil    
Path Found to a Combined MRI and CT Scanner
Wednesday, 20 March 2013   by spectrum.ieee.org    
Searchterm 'Field of View' was also found in the following services: 
spacer
Radiology  (10) Open this link in a new windowUltrasound  (4) Open this link in a new window
Rectangular Field of View
 
(RFOV) A different field of view (the scanned region) in the frequency and phase encoding directions that means the data acquisition with fewer measurement lines. Because there are fewer rows than columns, a rectangular image is obtained. To reduce the FOV in phase encoding direction (foldover direction) saves scan time by decreasing signal but invariable spatial resolution.
Also called HFI or undersampling.
mri safety guidance
Image Guidance
If the scanned object is oval, e.g. head or abdomen, a rectangular FOV is an easy to use scan parameter to reduce the scan time without loss of resolution.
spacer

• View the DATABASE results for 'Rectangular Field of View' (2).Open this link in a new window

MRI Resources 
Breast Implant - Corporations - Absorption and Emission - - MRI Centers - Safety Products
 
Aliasing ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this MRI artifact.
Artifact Information
NAME
Aliasing, backfolding, foldover, phase wrapping, wrap around
DESCRIPTION
Image wrap around
Aliasing is an artifact that occurs in MR images when the scanned body part is larger than field of view (FOV). As a consequence of the acquired k-space frequencies not being sampled densely enough, whereby portions of the object outside of the desired FOV get mapped to an incorrect location inside the FOV. The cyclical property of the Fourier transform fills the missing data of the right side with data from behind the FOV of the left side and vice versa. This is caused by a too small number of samples acquired in, e.g. the frequency encoding direction, therefore the spectrums will overlap, resulting in a replication of the object in the x direction.
Aliasing in the frequency direction can be eliminated by twice as fast sampling of the signal or by applying frequency specific filters to the received signal.
A similar problem occurs in the phase encoding direction, where the phases of signal-bearing tissues outside of the FOV in the y-direction are a replication of the phases that are encoded within the FOV. Phase encoding gradients are scaled for the field of view only, therefore tissues outside the FOV do not get properly phase encoded relative to their actual position and 'wraps' into the opposite side of the image.
mri safety guidance
Image Guidance
Use a larger FOV, RFOV or 3D Volume, apply presaturation pulses to the undesired tissue, adjust the position of the FOV, or select a small coil which will only receive signal from objects inside or near the coil. The number of phase encoding steps must be increased in phase direction, unfortunately resulting in longer scan times.
When this is not possible it can be corrected by oversampling the data. Aliasing is eliminated by Oversampling in frequency direction. No Phase Wrap (Foldover Suppression) options typically correct the phase encoding by doubling the field of view, doubling the number of phase encodes (to keep resolution constant) and halving the number of averages (to keep scan time constant) then discarding the additional data and processing the image within the desired field of view (but this is more time consuming).
Tissue outside this doubled area can be folded nevertheless into the image as phase wrap. In this case combine more than 2 number of excitations / number of signal averages with foldover suppression.
See also Aliasing, Foldover Suppression, Oversampling, and Artifact Reduction - Aliasing.
spacer

• View the DATABASE results for 'Aliasing Artifact' (11).Open this link in a new window

Searchterm 'Field of View' was also found in the following services: 
spacer
News  (6)  Resources  (2)  Forum  (3)  
 
Backfolding ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Backfolding, foldover, phase wrapping, wrap around
DESCRIPTION
Image wrap around
Backfolding always occurs due to wrong phase encoding caused by objects outside the planned FOV. Phase encoding gradients are scaled for the field of view only. Tissues outside the FOV do not get properly phase encoded relative to their actual position and 'wraps' into the opposite side of the image. The Backfolding artifact projects image contents which fall outside the imaging FOV back into the image; the back folded information thus reappearing on the other side of the image. In fact, information along the phase encoding direction can be viewed as projected onto a cylindrical screen with a circumference corresponding to the linear field of view dimension in the phase encoding direction.

See also Aliasing Artifact.
spacer

• View the DATABASE results for 'Backfolding Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
Searchterm 'Field of View' was also found in the following services: 
spacer
Radiology  (10) Open this link in a new windowUltrasound  (4) Open this link in a new window
Machine Imperfection ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Machine imperfection, data error
DESCRIPTION
Striped ghosts with a shift of half the field of view
REASON
Non-uniform sampling, phase differences
HELP
Data correction
Machine imperfection-based artifacts manifest themselves due to the fact that the odd k-space lines are acquired in a different direction than the even k-space lines. Slight differences in timing result in shifts of the echo in the acquisition window. By the shift theorem, such shifts in the time domain data then produce linear phase differences in the frequency domain data.
Without correction, such phase differences in every second line produce striped ghosts with a shift of half the field of view, so-called Nyquist ghosts. Shifts in the applied magnetic field can also produce similar (but constant in amplitude) ghosts.
This artifact is commonly seen in an EPI image and can arise from both, hardware and sample imperfections.
A further source of machine-based artifact arises from the need to acquire the signal as quickly as possible. For this reason the EPI signal is often acquired during times when the gradients are being switched. Such sampling effectively means that the k-space sampling is not uniform, resulting in ringing artifacts in the image.
mri safety guidance
Image Guidance
Such artifacts can be minimized by careful setup of the spectrometer and/or correction of the data. For this reasons reference data are often collected, either as a separate scan or embedded in the imaging data. The non-uniform sampling can be removed by knowing the form of the gradient switching. It is possible to regrid the data onto a uniform k-space grid.
spacer

• View the DATABASE results for 'Machine Imperfection Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
Pacemaker - Research Labs - MRA - MRI Technician and Technologist Career - Abdominal Imaging - Cardiovascular Imaging
 
     1 - 5 (of 29)     next
Result Pages : [1]  [2 3 4 5 6]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]