During frequency encoding, fat protons precess slower than water protons in the same slice because of their magnetic shielding. Through the difference in resonance frequency between water and fat, protons at the same location are misregistrated (dislocated) by the Fourier transformation, when converting MRI signals from frequency to spatial domain. This chemical shift misregistration cause accentuation of any fat-water interfaces along the frequency axis and may be mistaken for pathology. Where fat and water are in the same location, this artifact can be seen as a bright or dark band at the edge of the anatomy.
Protons in fat and water molecules are separated by a chemical shift of about 3.5 ppm. The actual shift in Hertz (Hz) depends on the magnetic field strength of the magnet being used. Higher field strength increases the misregistration, while in contrast a higher gradient strength has a positive effect. For a 0.3 T system operating at 12.8 MHz the shift will be 44.8 Hz compared with a 223.6 Hz shift for a 1.5 T system operating at 63.9 MHz.
The selective excitation of spins in only a limited region of space. This can be particularly useful for spectroscopy as well as imaging. Spatial localization of the signal source may be achieved through spatially selective excitation and the resulting signal may be analyzed directly for the spectrum corresponding to the excited region. It is usually achieved with selective excitation.
Typically, a single dimension of localization can be achieved with one selective RF excitation pulse (and a magnetic field gradient along a desired direction), while a localized volume (3D) can be excited with a stimulated echo produced with three selective RF pulses whose selective magnetic fieldgradients are mutually orthogonal, having a common intersection in the desired region. Similar 'crossed plane' excitation can be used with selective 180° refocusing pulses and conventional spinechoes.
A degree of spatial localization of excitation can alternatively be achieved with depth pulses, e.g. when using surface coils for excitation as well as signal detection. An indirect application of selective excitation for volume-selected spectroscopy is to use appropriate combinations of signals acquired after selective inversion of different regions, in order to subtract away the signal from undesired regions.