Chemical Shift ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Chemical shift, black boundary, spatial misregistration, relief
DESCRIPTION
Black or bright band
During frequency encoding, fat protons precess slower than water protons in the same slice because of their magnetic shielding. Through the difference in resonance frequency between water and fat, protons at the same location are misregistrated (dislocated) by the Fourier transformation, when converting MRI signals from frequency to spatial domain. This chemical shift misregistration cause accentuation of any fat-water interfaces along the frequency axis and may be mistaken for pathology. Where fat and water are in the same location, this artifact can be seen as a bright or dark band at the edge of the anatomy.
Protons in fat and water molecules are separated by a chemical shift of about 3.5 ppm. The actual shift in Hertz (Hz) depends on the magnetic field strength of the magnet being used. Higher field strength increases the misregistration, while in contrast a higher gradient strength has a positive effect. For a 0.3 T system operating at 12.8 MHz the shift will be 44.8 Hz compared with a 223.6 Hz shift for a 1.5 T system operating at 63.9 MHz.
mri safety guidance
Image Guidance
For artifact reduction helps a smaller water fat shift (higher bandwidth), a higher matrix, an in phase TE or a spin echo technique. Since the misregistration offset is present in the read out axis the patient may be rescanned with this axis parallel to the fat-water interface. Steeper gradient may be employed to reduce the chemical shift offset in mm. Another strategy is to employ specialized pulse sequences such as fat saturation or inversion recovery imaging. Fat suppression techniques eliminate chemical shift artifacts caused by the lack of fat signal.

See also Black Boundary Artifact and Magnetic Resonance Spectroscopy.
spacer
 
• Related Searches:
    • Matrix Size
    • Chemical Shift Imaging
    • Chemical Shift Spatial Offset
    • Bandwidth
    • Opposed Phase Image
 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
  News & More:
What is chemical shift artefact? Why does it occur? How many Hz at 1.5 T?
   by www.revisemri.com    
Abdominal MRI at 3.0 T: The Basics Revisited
Wednesday, 20 July 2005   by www.ajronline.org    
MRI Resources 
Movies - MRCP - Musculoskeletal and Joint MRI - Open Directory Project - MRI Training Courses - Functional MRI
 
Volume Selective Excitation
 
The selective excitation of spins in only a limited region of space. This can be particularly useful for spectroscopy as well as imaging. Spatial localization of the signal source may be achieved through spatially selective excitation and the resulting signal may be analyzed directly for the spectrum corresponding to the excited region. It is usually achieved with selective excitation.
Typically, a single dimension of localization can be achieved with one selective RF excitation pulse (and a magnetic field gradient along a desired direction), while a localized volume (3D) can be excited with a stimulated echo produced with three selective RF pulses whose selective magnetic field gradients are mutually orthogonal, having a common intersection in the desired region. Similar 'crossed plane' excitation can be used with selective 180° refocusing pulses and conventional spin echoes.
A degree of spatial localization of excitation can alternatively be achieved with depth pulses, e.g. when using surface coils for excitation as well as signal detection. An indirect application of selective excitation for volume-selected spectroscopy is to use appropriate combinations of signals acquired after selective inversion of different regions, in order to subtract away the signal from undesired regions.
spacer

• View the DATABASE results for 'Volume Selective Excitation' (3).Open this link in a new window

MRI Resources 
Safety Training - RIS - Societies - Pregnancy - Movies - Image Quality
 
previous      6 - 7 (of 7)     
Result Pages : [1]  [2]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]