Opposed Phase ImageInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
An image in which the signal from two spectral components (such as fat and water) is 180° out of phase and leads to destructive interference in a voxel.
Since fat precesses slower than water, based on their chemical shift, their signals will decay and precess in the transverse plane at different frequencies. When the phase of the TE becomes opposed (180°), their combined signal intensities subtract with each other in the same voxel, producing a signal void or dark band at the fat/water interface of the tissues being examined.
Opposed phase gradient echo imaging for the abdomen is a lipid-type tissue sensitive sequence particularly for the liver and adrenal glands, which puts a signal intensity around abnormal water-based tissues or lesions that are fatty. Due to the increased sensitivity of opposed phase, the tissue visualization increases the lesion-to-liver contrast and exhibits more signal intensity loss in tissues containing small amounts of lipids compared to a spin echo T1 with fat suppression. Using an opposed phase gradient echo also provides the ability to differentiate various pathologies in the brain, including lipids, methaemoglobin, protein, calcifications and melanin.

See also Out of Phase, and Dixon.
 
Images, Movies, Sliders:
 MRI Liver Out Of Phase  Open this link in a new window
    
 
spacer
 
• Related Searches:
    • Field Echo with Echo Time set for Water and Fat Signals in Opposition
    • Black Boundary Artifact
    • Chemical Shift Imaging
    • Chemical Shift Artifact
    • Chemical Shift
 
Further Reading:
  News & More:
Adrenal Myelolipoma
Tuesday, 19 June 2001   by www.emedicine.com    
Iron overload: accuracy of in-phase and out-of-phase MRI as a quick method to evaluate liver iron load in haematological malignancies and chronic liver disease
Friday, 1 June 2012   by www.ncbi.nlm.nih.gov    
Searchterm 'Interference' was also found in the following services: 
spacer
Radiology  (2) Open this link in a new windowUltrasound  (10) Open this link in a new window
Penetration Panel
 
This is the point of penetration by the MR system cables into the RF room. This panel contains filters and other connections required to provide communication between the scan room and the rest of the system without allowing interference to pass.
spacer
MRI Resources 
Shielding - Examinations - Cardiovascular Imaging - Directories - Claustrophobia - Nerve Stimulator
 
Unshielded Electric Components
 
Unshielded electric components in the magnet room will excite RF noise artifacts (interference) in different variations. RF 'buzz' will be seen over a range of frequencies on the images. The artifact runs in frequency encoding direction.
mri safety guidance
Image Guidance
Take care to use only MR compatible (special shielded) devices.

See also MR Compatibility.
spacer
 
Further Reading:
  News & More:
Electrical Noise in Intraoperative Magnetic Resonance Imaging reveals the intricacies of human brain – integrated consciousness with connectors to chilled universe
Sunday, 18 January 2009   by www.indiadaily.com    
Searchterm 'Interference' was also found in the following services: 
spacer
News  (3)  Resources  (3)  Forum  (2)  
 
Zipper ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Zipper, star
DESCRIPTION
Bands through image center
REASON
Hardware or software problems
HELP
Larger FOV, oversampling
Zipper artifacts appear as dashed lines. There are various causes for this MRI artifact.
Most of zipper artifacts result from inhomogeneities of the magnetic field caused by interferences with radio frequency from various sources ('your radio is working in the scanner room means your shielding is not working'). Software and equipment problems can also cause zipper lines in both directions.
mri safety guidance
Image Guidance
Scanning room door during the acquisition of images is closed, shielding is working, and all disturbing devices are removed?

See also Room Shielding, Active Shielding, Radio Frequency Shielding.
spacer
Searchterm 'Interference' was also found in the following services: 
spacer
Radiology  (2) Open this link in a new windowUltrasound  (10) Open this link in a new window
previous      11 - 14 (of 14)     
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]