 MZ            Longitudinal Magnetization 
Searchterm 'MZ' was found in the Abbreviation Register. 
Result : Searchterm 'MZ' found in 1 term [] and 6 definitions []
1 - 5 (of 7)     next
Result Pages : [1]  [2]
MRI Resources 
Open Directory Project - Most Wanted - Anatomy - RIS - Societies - NMR
 
Mz
 
(MZ) Dimension in the direction of the static magnetic field (B0 and H0), in both the stationary and rotating frames of reference.
spacer
 
• Share the entry 'Mz':  Facebook  Twitter  LinkedIn  
MRI Resources 
Sequences - Anatomy - Equipment - Nerve Stimulator - Safety Products - Developers
 
Spin Lattice Relaxation Time
 
(T1) The spin lattice relaxation time (also called longitudinal relaxation time and T1 Time) is a spin property, whereby the value changes between different tissues. By the spin lattice relaxation process, the longitudinal magnetization Mz achieve the equilibrium value Mz0. The T1 time constant is an exponential approach toward Mz0.
The equation for the magnetization at a time t will be (if at t=0 the longitudinal magnetization is Mz0):
Mz(t) = M0+(Mz (0) - Mz0) exp(t/T1)
spacer

• View the DATABASE results for 'Spin Lattice Relaxation Time' (2).Open this link in a new window

 
Further Reading:
  Basics:
Electron Spin Resonance
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
MRI's inside story
Thursday, 4 December 2003   by www.economist.com    
MULTIEXPONENTIAL PROTON SPIN-SPIN RELAXATION IN MAGNETIC RESONANCE IMAGING OF HUMAN BRAIN TUMORS
Friday, 26 March 1999   by www.dkfz-heidelberg.de    
MRI Resources 
Distributors - Safety Products - Portals - Services and Supplies - Corporations - Equipment
 
Inversion Recovery SequenceForum -
related threadsInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
Inversion Recovery Sequence Timing Diagram (IR) The inversion recovery pulse sequence produces signals, which represent the longitudinal magnetization existing after the application of a 180° radio frequency pulse that rotates the magnetization Mz into the negative plane. After an inversion time (TI - time between the starting 180° pulse and the following 90° pulse), a further 90° RF pulse tilts some or all of the z-magnetization into the xy-plane, where the signal is usually rephased with a 180° pulse as in the spin echo sequence. During the initial time period, various tissues relax with their intrinsic T1 relaxation time.
In the pulse sequence timing diagram, the basic inversion recovery sequence is illustrated. The 180° inversion pulse is attached prior to the 90° excitation pulse of a spin echo acquisition. See also the Pulse Sequence Timing Diagram. There you will find a description of the components.
The inversion recovery sequence has the advantage, that it can provide very strong contrast between tissues having different T1 relaxation times or to suppress tissues like fluid or fat. But the disadvantage is, that the additional inversion radio frequency RF pulse makes this sequence less time efficient than the other pulse sequences.

Contrast values:
PD weighted: TE: 10-20 ms, TR: 2000 ms, TI: 1800 ms
T1 weighted: TE: 10-20 ms, TR: 2000 ms, TI: 400-800 ms
T2 weighted: TE: 70 ms, TR: 2000 ms, TI: 400-800 ms

See also Inversion Recovery, Short T1 Inversion Recovery, Fluid Attenuation Inversion Recovery, and Acronyms for 'Inversion Recovery Sequence' from different manufacturers.
 
Images, Movies, Sliders:
 Brain MRI Inversion Recovery  Open this link in a new window
    
 Knee MRI Sagittal STIR 002  Open this link in a new window
 Brain MRI Coronal FLAIR 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Inversion Recovery Sequence' (8).Open this link in a new window

 
Further Reading:
  Basics:
The equation for a repeated inversion recovery sequence
Contrast mechanisms in magnetic resonance imaging
2004   by www.iop.org    
  News & More:
FLAIR Vascular Hyperintensity: An Important MRI Marker in Patients with Transient Ischemic Attack
Thursday, 14 July 2022   by www.dovepress.com    
MRI Resources 
Pacemaker - Quality Advice - Brain MRI - MRI Training Courses - MRCP - MRI Technician and Technologist Career
 
Longitudinal Magnetization
 
(Mz) The component of the net magnetization vector in the direction of the static magnetic field (z). After RF excitation, this vector returns to its equilibrium value at a rate characterized by the time constant T1.
spacer

• View the DATABASE results for 'Longitudinal Magnetization' (25).Open this link in a new window

MRI Resources 
Examinations - Crystallography - Societies - MRI Physics - Shielding - Diffusion Weighted Imaging
 
Signal Intensity
 
Signal intensity interpretation in MR imaging has a major problem.
Often there is no intuitive approach to signal behavior as signal intensity is a very complicated function of the contrast-determining tissue parameter, proton density, T1 and T2, and the machine parameters TR and TE. For this reason, the terms T1 weighted image, T2 weighted image and proton density weighted image were introduced into clinical MR imaging.
Air and bone produce low-intensity, weaker signals with darker images. Fat and marrow produce high-intensity signals with brighter images.
The signal intensity measured is related to the square of the xy-magnetization, which in a SE pulse sequence is given by
Mxy = Mxy0(1-exp(-TR/T1)) exp(-TE/T2) (1)
where Mxy0 = Mz0 is proportional to the proton or spin density, and corresponds to the z-magnetization present at zero time of the experiment when it is tilted into the xy-plane.

See also T2 Weighted Image and Ernst Angle.
spacer

• View the DATABASE results for 'Signal Intensity' (56).Open this link in a new window


• View the NEWS results for 'Signal Intensity' (1).Open this link in a new window.
 
Further Reading:
  Basics:
Contrast mechanisms in magnetic resonance imaging
2004   by www.iop.org    
Image Characteristics and Quality
   by www.sprawls.org    
  News & More:
Specialized MRI sensor can detect light deep within tissues
Thursday, 22 December 2022
Study compares effect of food intake on manganese-based MRI contrast agent absorption
Saturday, 3 December 2022   by www.itnonline.com    
Kidney stone imaging with 3D ultra-short echo time (UTE) magnetic resonance imaging. A phantom study - Abstract
Wednesday, 11 March 2015   by www.urotoday.com    
MRI Resources 
Corporations - MR Myelography - Movies - RIS - Veterinary MRI - Manufacturers
 
     1 - 5 (of 7)     next
Result Pages : [1]  [2]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]