Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Partial Echo' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Partial Echo' found in 1 term [] and 3 definitions [], (+ 11 Boolean[] results
1 - 5 (of 15)     next
Result Pages : [1]  [2 3]
MRI Resources 
MRI Reimbursement - Musculoskeletal and Joint MRI - Coils - Pregnancy - Safety Products - MRI Accidents
 
Partial Echo
 
(PE) The partial echo technique (also called fractional echo) is used to shorten the minimum echo time. By the acquisition of only a part of k-space data this technique benefits (like all partial Fourier techniques) from the complex conjugate symmetry between the k-space halves (this is called Hermitian symmetry).
The dephasing gradient in the frequency direction is reduced, and the duration of the readout gradient and the data acquisition window are shortened. Partial echo gives a better SNR at a given TE when a smaller FOV or thinner slices are selected, allows a longer sampling time, and a larger water fat shift (WFS, see also bandwidth) due to a lower gradient amplitude. The resolution is not affected. This is often used in gradient echo sequences (e.g. FLASH, Contrast Enhanced Magnetic Resonance Angiography) to reduce the echo time and yields a lower gradient moment. The disadvantage of using a partial echo can be a lower SNR, although this may be partly offset by the reduced echo time.
Also called Fractional Echo, Read Conjugate Symmetry, Single Side View.

See also Partial Fourier Technique and acronyms for 'partial echo' from different manufacturers.
spacer
 
• Share the entry 'Partial Echo':  Facebook  Twitter  LinkedIn  
 
Further Reading:
  Basics:
Method and apparatus for subterranean formation flow imaging
   by www.google.com    
MRI Resources 
Jobs pool - Brain MRI - Databases - Anatomy - Quality Advice - Portals
 
Fractional Echo
 
Fractional echo (also called asymmetric or partial echo) is used to shorten the echo time in a sequence, by acquiring partial echoes in the frequency direction. The reduction of echo time is possible because if the first part of the echo is not received, the dephasing lobe of the frequency encoding gradient is not to be on for quite as long, and this saves time.

See also Partial Fourier Technique, Read Conjugate Symmetry, Single Side View and acronyms for 'fractional echo' from different manufacturers.
spacer

• View the DATABASE results for 'Fractional Echo' (2).Open this link in a new window

 
Further Reading:
  News & More:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
MRI Resources 
Non-English - Collections - Societies - Jobs - Chemistry - Implant and Prosthesis pool
 
Read Conjugate SymmetryInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
This function is similar to partial averaging in that the complex conjugate relationship between the both halves of k-space is used to shorten the echo time in an imaging sequence.

See Partial Echo.
spacer

• View the DATABASE results for 'Read Conjugate Symmetry' (3).Open this link in a new window

MRI Resources 
Spine MRI - Anatomy - DICOM - Liver Imaging - Pregnancy - Services and Supplies
 
Single Side ViewInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
spacer

• View the DATABASE results for 'Single Side View' (3).Open this link in a new window

MRI Resources 
Anatomy - Artifacts - Hospitals - IR - DICOM - Non-English
 
Partial Saturation Spin EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PSSE) Partial saturation sequence in which the signal is detected as a spin echo. Even though a spin echo is used, there will not necessarily be a significant contribution of the T2 relaxation time to image contrast, unless the echo time, TE, is on the order of or longer than T2.
spacer
MRI Resources 
Intraoperative MRI - Mass Spectrometry - Collections - Implant and Prosthesis - MRI Training Courses - DICOM
 
     1 - 5 (of 15)     next
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]