Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Relaxation Time' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Relaxation Time' found in 5 terms [] and 52 definitions []
previous     26 - 30 (of 57)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12]
Searchterm 'Relaxation Time' was also found in the following services: 
spacer
News  (3)  Resources  (2)  Forum  (6)  
 
Flow Effects
 
Motion of material being imaged, particularly flowing blood, can result in many possible effects in the images.
Fast moving blood produces flow voids, blood flowing in to the outer slices of an imaging volume produces high signals (flow related enhancement, entry slice phenomenon), pulsatile flow creates ghost images of the vessel extending across the image in the phase encoding direction (image misregistration).
Flow-related dephasing occurring when spin isochromats are moving with different velocities in an external gradient field G so that they acquire different phases. When these phases vary by more then 180° within a voxel, substantial spin dephasing results leading to considerable intravascular signal loss.
These effects can be understood as caused by time of flight effects (washout or washin due to motion of nuclei between two consecutive spatially selective RF excitations, repeated in times on the order of, or shorter than the relaxation times of blood) or phase shifts (delay between phase encoding and frequency encoding) that can be acquired by excited spins moving along magnetic field gradients.
The inconsistency of the signal resulting from pulsatile flow can lead to artifacts in the image. The flow effects can also be exploited for MR angiography or flow measurements.

See also Flow Artifact.
 
Images, Movies, Sliders:
 Anatomic MRI of the Knee 1  Open this link in a new window
    
SlidersSliders Overview

 Anatomic MRI of the Neck  Open this link in a new window
    
SlidersSliders Overview

 PCA-MRA 3D Brain Venography Colored MIP  Open this link in a new window
    

 TOF-MRA Circle of Willis Inverted MIP  Open this link in a new window
    

 
spacer
 
Further Reading:
  News & More:
Magnetic resonance flow velocity and temperature mapping of a shape memory polymer foam device
Thursday, 31 December 2009   by 7thspace.com    
MRI measure of blood flow over atherosclerotic plaque may detect dangerous plaque
Friday, 5 April 2013   by www.sciencecodex.com    
MRI Resources 
Spectroscopy pool - Cochlear Implant - Jobs pool - MRA - Stent - Cardiovascular Imaging
 
Fluid Attenuation Inversion RecoveryInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(FLAIR) Fluid attenuation inversion recovery is a special inversion recovery sequence with long TI to remove the effects of fluid from the resulting images. The TI time of the FLAIR pulse sequence is adjusted to the relaxation time of the component that should be suppressed. For fluid suppression the inversion time (long TI) is set to the zero crossing point of fluid, resulting in the signal being 'erased'.
Lesions that are normally covered by bright fluid signals using conventional T2 contrast are made visible by the dark fluid technique FLAIR is an important technique for the differentiation of brain and spine lesions.

See also Inversion Recovery.
 
Images, Movies, Sliders:
 Brain MRI Coronal FLAIR 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Fluid Attenuation Inversion Recovery' (5).Open this link in a new window

 
Further Reading:
  Basics:
Newer Sequences for Spinal MR Imaging: Smorgasbord or Succotash of Acronyms?
   by www.ajnr.org    
  News & More:
What MRI-Derived Data and Other Factors Reveal About White Matter Hyperintensity in Former Football Players
Saturday, 23 December 2023   by www.diagnosticimaging.com    
FLAIR Vascular Hyperintensity: An Important MRI Marker in Patients with Transient Ischemic Attack
Thursday, 14 July 2022   by www.dovepress.com    
Early Identification of Ischemic Stroke With DWI-FLAIR Mismatch
Wednesday, 5 January 2011   by www.doctorslounge.com    
MRI Resources 
Spine MRI - Liver Imaging - Contrast Agents - Mass Spectrometry - Developers - Absorption and Emission
 
Free Induction Decay
 
(FID) A free induction decay curve is generated as excited nuclei relax. The amplitude of the FID signal becomes smaller over time as net magnetization returns to equilibrium. If transverse magnetization of the spins is produced, e.g. by a 90° pulse, a transient MR signal will result that will decay toward zero with a characteristic time constant T2 (or T2*); this decaying signal is the free induction decay.
The signal peaks of the echoes fall onto this T2 decay curve, while at each echo the signals arise and decay with T2*. The typical T2 relaxation times being of the order of 5-200 ms in the human body. The first part of the FID is not observable (named the 'receiver dead time') caused by residual effects of the powerful exciting radio frequency pulse on the electronics of the receiver.
spacer

• View the DATABASE results for 'Free Induction Decay' (8).Open this link in a new window

 
Further Reading:
  Basics:
Free induction decay
   by en.wikipedia.org    
  News & More:
Magnetic resonance imaging
   by www.scholarpedia.org    
Searchterm 'Relaxation Time' was also found in the following services: 
spacer
News  (3)  Resources  (2)  Forum  (6)  
 
Gadolinium OxideInfoSheet: - Contrast Agents - 
Intro, Overview, 
Characteristics, 
Types of, 
etc.
 
Gadolinium oxide (Gd2O3) is a prototype paramagnetic agent for contrast enhanced MRI. Gd2O3 particles have very high relaxivity. With its high magnetic moment, gadolinium reduces the relaxation time and enhance MR signal intensity. All gadolinium compounds are highly toxic. Very stable complexes are developed to eliminate the toxicity. See also Paramagnetic Substance, Paramagnetism, Paramagnetic Contrast Agents and Contrast Agents, the info sheet gives an overview and more in-dept information about different types of MRI Contrast media.
spacer

• View the DATABASE results for 'Gadolinium Oxide' (2).Open this link in a new window

 
Further Reading:
  Basics:
Gadolinium oxide nanoparticles enhance MRI contrast
Thursday, 29 September 2011   by nanotechweb.org    
  News & More:
Northern Rare Earth eyes medical device prospects
Thursday, 29 December 2016   by usa.chinadaily.com.cn    
MRI Resources 
IR - Claustrophobia - Developers - Process Analysis - Stimulator pool - Safety pool
 
Gastrointestinal Paramagnetic Contrast AgentsInfoSheet: - Contrast Agents - 
Intro, Overview, 
Characteristics, 
Types of, 
etc.MRI Resource Directory:
 - Contrast Agents -
 
Paramagnetic substances, for example Gd-DTPA solutions, are used as MRI oral contrast agents in gastrointestinal imaging to depict the lumen of the digestive organs. Different Gd-DTPA solutions or zeolites containing gadolinium can be used e.g., for diagnosis of delayed gastric emptying, diagnosis of Crohn's disease etc.
Low concentrations of gastrointestinal paramagnetic contrast agents cause a reduction in T1 relaxation time; consequently, these agents act on T1 weighted images by increasing the signal intensity of the bowel lumen. High concentrations cause T2 shortening by decreasing the signal, similar to superparamagnetic iron oxide. Gd-DTPA chelates are unstable at the low pH in the stomach, therefore buffering is necessary for oral use.

See also Gadopentetate Gastrointestinal, Gadolinium Zeolite, Negative Oral Contrast Agents, Gastrointestinal Superparamagnetic Contrast Agents, and Ferric ammonium citrate.
 
Images, Movies, Sliders:
 MR Colonography Gadolinium per Rectum  Open this link in a new window
      

Courtesy of  Robert R. Edelman
 
spacer

• View the DATABASE results for 'Gastrointestinal Paramagnetic Contrast Agents' (5).Open this link in a new window

MRI Resources 
RIS - MRA - Spectroscopy - Corporations - Universities - MRCP
 
previous      26 - 30 (of 57)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]