Passive Shielding
 
Magnetic shielding through the use of high permeability material. The iron provides a return path for the stray field lines of magnetic flux and so significantly decreases the flux away from the magnet.
Passive shielding (see also Faraday cage) significantly eases the problems of siting a MR imager in a confined space. Ferromagnetic objects are less prone to being attracted to the magnet, ancillary electronic equipment, credit cards and computer disks can be brought closer to the magnet and the MRI safety limit for pacemaker wearers (the 5 gauss line = 0.5 mT) is reduced from, typically, 10 m to 2 m from the magnet. A passive shield for a whole-body MRI magnet weights many tons. An alternative method of controlling stray field is active shielding.

See also Active Shielding, Magnetic Shielding, Self Shielding and Room Shielding.
spacer
 
• Related Searches:
    • Active Shielding
    • Magnetic Shielding
    • Self Shielding
    • Faraday Cage
    • MRI Safety
 
Further Reading:
  Basics:
Faraday's Law
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
Magnetic Sensitivity of MRI Systems to External Iron: The Design Process
   by www.integratedsoft.com    
MRI Resources 
Knee MRI - Quality Advice - Sequences - Functional MRI - General - Cardiovascular Imaging
 
Phase 1, 2, 3, 4 Drug Trials
 
Different stages of testing drugs in humans, from first application in humans through limited and broad clinical tests, to postmarketing studies. Preclinical trials are the testing in animals.
Phase I: Safety, pharmacokinetics
Phase II: Dose
Phase III: Efficacy
Phase IV: Postmarketing
spacer
MRI Resources 
Liver Imaging - Bioinformatics - MRCP - Patient Information - Hospitals - Artifacts
 
Quench
 
Quenching is the cryogen boil off (liquid helium) which is used to cool the superconducting magnet coils of high field MRI systems. This results in a loss of superconductivity in the magnet, in a rapid increase in the resistivity of the magnet, which generates heat that results in further evaporation of the cryogen.
mri safety guidance
MRI Safety Guidance
A quench can cause total magnet failure and should be avoided.
spacer

• View the DATABASE results for 'Quench' (5).Open this link in a new window

 
Further Reading:
  Basics:
Cryogenic Liquids and their Hazards
   by www.ccohs.ca    
Searchterm 'Safety' was also found in the following services: 
spacer
News  (69)  Resources  (43)  Forum  (15)  
 
Radio FrequencyForum -
related threads
 
(RF) Radio frequency refers to that portion of the electromagnetic spectrum in which electromagnetic waves can be generated by alternating current fed to an antenna.
The RF pulses used in MRI are commonly in the 1-100 megahertz range, and their effect upon a body is potential heating of tissues and foreign bodies, such as metallic implants, mainly at the surface.
This is a principal area of concern for MRI safety limits caused by absorption (see specific absorption rate) of the applied pulses of RF energy.
mri safety guidance
Image Guidance
The higher the frequency, the larger will be the amount of heat developed.
The more ionic the biochemical environment in the tissue, the more energy that will be deposited as heat.
This effect is well known for homogeneous model systems, but the complex structure of various human tissues makes detailed theoretical calculations very difficult, if not impossible. By scanning problems, it is important to verify the transmission frequency. If the RF transmitted into the patient was, e.g. 5000 Hz lower than the resonance frequency of the protons, no protons was excited, and no signal returns.
spacer

• View the DATABASE results for 'Radio Frequency' (72).Open this link in a new window


• View the NEWS results for 'Radio Frequency' (2).Open this link in a new window.
 
Further Reading:
  Basics:
Radio frequency
   by en.wikipedia.org    
  News & More:
On the estimation of the worst-case implant-induced RF-heating in multi-channel MRI.
Thursday, 2 March 2017   by www.ncbi.nlm.nih.gov    
Safety and reliability of Radio Frequency Identification Devices in Magnetic Resonance Imaging and Computed Tomography
Thursday, 4 February 2010   by 7thspace.com    
Radiologists Offer Non-surgical Treatment For Early-stage Liver Cancer
Saturday, 5 March 2005   by www.sciencedaily.com    
MRI Resources 
Safety pool - Cardiovascular Imaging - Directories - Non-English - Manufacturers - MRI Physics
 
Radio Frequency Shielding
 
Radio frequency shielding includes the construction of enclosures for the purpose of reducing the transmission of electric or magnetic fields from one space to another (Faraday cage, Faraday shield). Electrically conducted shielding is designed to isolate MRI systems from its environment at the resonant frequencies.
All electronic and computer systems radiate certain frequencies of radio and magnetic waves. They can interfere with other equipment in the vicinity. Magnetic shielding enclosures are used to reduce the levels of RF radiation that enters or leaves the shielded room.
Copper shielding enclosures are designed to filter a range of frequencies under specified conditions. One of the characteristics of copper is its high electrical conductivity. Also its other physical properties like ductility, malleability, and ease of soldering, make it an ideal material for radio frequency shielding. Sheet copper can be formed into any shape and size, and electrically connected to a grounding system to provide an effective RF shielding.

See also MRI Safety
spacer

• View the DATABASE results for 'Radio Frequency Shielding' (3).Open this link in a new window

MRI Resources 
Knee MRI - Most Wanted - Spectroscopy pool - Absorption and Emission - Safety pool - MRI Technician and Technologist Career
 
previous      46 - 50 (of 57)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]