Signa Contour/i 0.5T™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gemedical.co.kr/rad/mri/products/contour/contouri.html From GE Healthcare;
GE's Signa Contour/i system uses the innovations like K4 technology and real-time interactive imaging. This compact magnet with wide-flare gantry obtains high patient comfort with low costs.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Compact
Head and body coil standard; all other coils optional; open architecture makes system compatible with a wide selection of coils
SYNCHRONIZATION
ECG/peripheral, respiratory gating
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, Angiography;; 2D/3D TOF, 2D/3D Phase Contrast;; 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, optional: EPI, 2D/3D Fiesta, FGRET, Spiral
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab
1 cm to 48 cm continuous
2D 0.8 mm to 20 mm; 3D 0.1 mm to 5 mm
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
POWER REQUIREMENTS
480 or 380/415 V
COOLING SYSTEM TYPE
Closed-loop water-cooled gradient
STRENGTH
SmartSpeed 23 mT/m, HiSpeed Plus 33 mT/m
Active
spacer
MRI Resources 
Claustrophobia - MRA - Implant and Prosthesis pool - Guidance - Used and Refurbished MRI Equipment - Raman Spectroscopy
 
Number of Signal Averages
 
(NSA) The average number for each phase encoding step. The NSA determine each distinct position-encoded signal to be used in image reconstruction.

See also Signal Averaging.
spacer

• View the DATABASE results for 'Number of Signal Averages' (5).Open this link in a new window

 
Further Reading:
  News & More:
CLINICAL WHOLE BODY MRI AT 3.0 T(.pdf)
2001
MRI Resources 
Supplies - MRI Training Courses - MR Guided Interventions - - Guidance - Services and Supplies
 
Nuclear Magnetic Resonance Signal
 
The electromagnetic signal in the radio-frequency range produced by the precession of the transverse magnetization of the spins. The rotation of the transverse magnetization induces a voltage in a receiving antenna (coil), which is amplified and demodulated by the receiver circuits. Electromagnetic signal in the radio frequency range produced by the precession of the transverse magnetization of the spins. The rotation of the transverse magnetization induces a voltage in a coil, which is amplified and demodulated by the receiver;; the signal may refer only to this induced voltage.
spacer

• View the DATABASE results for 'Nuclear Magnetic Resonance Signal' (2).Open this link in a new window

 
Further Reading:
  Basics:
Spin echoes, CPMG and T2 relaxation - Introductory NMR & MRI from Magritek
2013   by www.azom.com    
NMR Spectroscopy - Theory
   by www.shu.ac.uk    
  News & More:
A powder to enhance NMR signals
Thursday, 12 December 2013   by phys.org    
Searchterm 'Signa' was also found in the following services: 
spacer
News  (60)  Resources  (16)  Forum  (78)  
 
Signa Infinity 1.5T™ with ExciteInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/s_excite15/index.html From GE Healthcare;
'EXCITE technology has the potential to open the door to new imaging techniques and clinical applications, leaping beyond conventional two and three-dimensional MRI to true 4D imaging that will improve the diagnosis of disease throughout the human body from head to foot.' Robert R. Edelman, M.D., Professor of Radiology at Northwestern University Medical School and Chairman, Department of Radiology, at Evanston Northwestern Healthcare.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
Head and body coil standard; all other coils optional; open architecture makes system compatible with a wide selection of coils
Optional 2D/3D brain and prostate
SYNCHRONIZATION
ECG/peripheral, respiratory gating, (SmartPrep, SmartStep)
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, Angiography: 2D/3D TOF, 2D/3D Phase Contrast;; 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, EPI, optional: 2D/3D Fiesta, FGRET, Spiral, Tensor
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine
TR
1.3 to 12000 msec in increments of 1 msec
TE
0.4 to 2000 msec in increments of 1 msec
SINGLE/MULTI SLICE
Simultaneous scan and reconstruction;; up to 200 or 400 images per second
1 cm to 48 cm continuous
2D 0.7 mm to 20 mm; 3D 0.1 mm to 5 mm
1028 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
0.08 mm; 0.02 mm optional
MAGNET WEIGHT
3863 kg
H*W*D
172 x 208 x 216 cm
POWER REQUIREMENTS
480 or 380/415
COOLING SYSTEM TYPE
Closed-loop water-cooled gradient
CRYOGEN USE, L/hr
less than 0.03 L/hr liquid helium
STRENGTH
SmartSpeed 23 mT/m, HiSpeed Plus 33 mT/m, EchoSpeed Plus 33 mT/m
5-GAUSS FRINGE FIELD
4.0 m x 2.8 m axial x radial
Active
spacer

• View the DATABASE results for 'Signa Infinity 1.5T™ with Excite' (2).Open this link in a new window

MRI Resources 
Safety pool - Universities - - Fluorescence - Contrast Agents - Developers
 
Signa SP 0.5T™ Open ConfigurationInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/index.html From GE Healthcare;
The Signa SP 0.5T™ is an open MRI magnet that is designed for use in interventional radiology and intra-operative imaging. The vertical gap configuration increases patient positioning options, improves patient observation, and allows continuous access to the patient during imaging. The magnet enclosure also incorporates an intercom, patient observation video camera, laser patient alignment lights, and task lighting in the imaging volume.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Open
Integrated transmit and receive body coil; optional rotational body coil, head; other coils optional; open architecture makes system compatible with a wide selection of coilsarray
SYNCHRONIZATION
Standard cardiac gating, ECG/peripheral, respiratory gating
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, 2D/3D TOF, 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, EPI, optional: 2D/3D Fiesta, true chem sat, fat/water separation, single shot diffusion EPI
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine, slice and frequency zip, extended dynamic range, tailored RF
TR
1.3 to 12000 msec in increments of 1 msec
TE
0.4 to 2000 msec in increments of 1 msec
3cm to 40 cm continuous
2D: 1.4mm - 20mm 3D: 0.2mm - 20mm
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase//freq.
PIXEL INTENSITY
256 gray levels
60 x 58 cm
POWER REQUIREMENTS
200 - 480, 3-phase
COOLING SYSTEM TYPE
Liquid helium
0.03 L/hr, holds 300 L
STRENGTH
12 mT/m
Active
spacer

• View the DATABASE results for 'Signa SP 0.5T™ Open Configuration' (2).Open this link in a new window

 
Further Reading:
  News & More:
MR Surgical Suite, Improving surgical procedure quality (.pdf)
   by www3.gehealthcare.com    
MRI Resources 
Case Studies - Guidance - MRI Reimbursement - Mass Spectrometry - Open Directory Project - Used and Refurbished MRI Equipment
 
previous      16 - 20 (of 381)     next
Result Pages : [1 2 3 4 5]  [6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]