Slice Overlap ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Slice overlap
DESCRIPTION
Loss of signal
REASON
HELP
Overlapping prevention
The slice overlap artifact is another name for crosstalk artifact. If slices of multislice acquisitions are overlapping, the spinning nuclei belonging to more than one slice getting multiple times saturated, which leads to signal loss in this areas.
mri safety guidance
Image Guidance
This problem occurs often in cervical or lumbar spine MRI, when scanning each disc with multi angle oblique technique. If prevention of overlapping is not possible, try to position the saturated region posterior to the spinal canal, outside the region of interest.

See also Crosstalk (Crosstalk), and Multiple Slice Imaging.
spacer
 
Further Reading:
  Basics:
Slice-overlap Artifacts
   by www.mritutor.org    
Searchterm 'Slice' was also found in the following services: 
spacer
Radiology  (24) Open this link in a new windowUltrasound  (9) Open this link in a new window
Spacing between SlicesInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
The not measured distance between the slices. For some sequences there is a slightest space of e.g. 10-20% necessary. A lower spacing can introduce more noise into adjoining slices.
spacer
MRI Resources 
Universities - Quality Advice - Shoulder MRI - Breast Implant - Raman Spectroscopy - Online Books
 
Gradient Echo Multi SliceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(GEMS) This pulse sequence uses a changeable flip angle instead of a 90° pulse and a gradient instead of a RF pulse to rephase the FID.
T2*, T1 weighted and proton density images can be acquired. The flip angle in combination with the TR determines the T1 weighting and the TE controls the amount of dephasing. To minimize T2* the echo time should be short.

See also Gradient Echo Sequence.
spacer
Searchterm 'Slice' was also found in the following services: 
spacer
News  (6)  Resources  (5)  Forum  (25)  
 
Entry Slice Phenomenon (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Entry slice phenomenon
DESCRIPTION
Bright signals in blood vessels at the first slice
REASON
Unsaturated spins
The entry slice phenomenon arise in MRI when blood with unsaturated spins flows in the observed slice(s). These spins will emit a strong signal, because of their unsaturated status (flow related enhancement). The number of slices affected depends on the flow velocity and the slice thickness; the direction of flow determines which slices are affected. Time of Flight MRA is based on this entry slice phenomenon.

See also Flow Compensation, Flow Related Enhancement, Artifact Overview and Artifacts Reduction Index.
 
Images, Movies, Sliders:
 TOF-MRA Circle of Willis Inverted MIP  Open this link in a new window
    

 
spacer
 
Further Reading:
  News & More:
Troubleshooting the ACR MRI Accreditation Phantom Tests
   by www.aapm.org    
Searchterm 'Slice' was also found in the following services: 
spacer
Radiology  (24) Open this link in a new windowUltrasound  (9) Open this link in a new window
Contiguous Slice Multi Echo MultiplanarInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(CSMEMP) MR imaging pulse sequence in which signals for multiple contiguous slices at multiple echoes are collected.
spacer
MRI Resources 
Abdominal Imaging - Cochlear Implant - Hospitals - MRI Accidents - Coils - Nerve Stimulator
 
previous      11 - 15 (of 178)     next
Result Pages : [1 2 3 4]  [5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]