Relaxation Effect
 
The relaxation effect is the transition of an atom or molecule from a higher energy level to a lower one. The return of the excited proton from the high energy to the low energy level is associated with the loss of energy to the surrounding tissue. The T1 and T2 relaxation times define the way that the protons return to their resting levels after the initial radio frequency (RF) pulse. The T1 and T2 relaxation rates have an effect of the signal to noise ratio (SNR) of MR images.
The relaxation process is a result of both T1 and T2, and can be controlled by the dependency of one of the two biological parameters T1 and T2 in the recorded signal. A T1 weighted spin echo sequence is based on a short repetition time (TR) and a change of it will affect the acquisition time and the T1 weighting of the image. Increased TR results in improved SNR caused by longer recovering time for the longitudinal magnetization. Increased TE improves the T2 weighting, combined with a long TR (of several T1 times) to minimize the T1 effect.
spacer
 
Further Reading:
  News & More:
MRI's inside story
Thursday, 4 December 2003   by www.economist.com    
MRI Resources 
Portals - Education pool - Stimulator pool - Safety pool - Safety Training - Directories
 
Reverse Fast Imaging with Steady State PrecessionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(PSIF) A heavily T2* weighted contrast enhanced gradient echo (mirrored FISP) technique. Because TE is relatively long, there are much flow artifacts and less signal to noise. In normal gradient echo techniques a FID-signal results after the RF pulses. This FID is rephased very fast and just before the next FID follows a spin echo signal. The SE is spoiled in FLASH sequences, but with PSIF sequences, only the SE is measured, not the FID.
spacer

• View the DATABASE results for 'Reverse Fast Imaging with Steady State Precession' (2).Open this link in a new window

 
Further Reading:
  News & More:
Fast T2 weighted imaging by PSIF at 0.2T for interventional MRI.(.pdf)
   by cds.ismrm.org    
MRI Resources 
DICOM - Spine MRI - Safety Products - MRI Reimbursement - Nerve Stimulator - Colonography
 
Saturation
 
In MR, saturation is a nonequilibrium state with no net magnetization. The same amount of nuclear spins is aligned against and with the magnetic field. Saturation methods like FatSat, SPIR etc., work with a frequency selective saturation pulse for a specific chemical shift applied before the actual sequence starts. This saturation pulse adjusts the magnetization from tissue components to zero. The hydrogen nuclei of fat and water resonate at different frequencies, which makes it possible to excite just the fat with repeatedly applying RF pulses at the Larmor frequency with interpulse times compared to T1. The resulting signal is then destroyed with a gradient pulse (Spoiler Gradient Pulse). Fat is the chemical compound to be saturated at a fat saturation sequence. When the actual sequence follows, (e.g., a spin echo sequence) the unwanted suppressed component will not resonate.

See also Saturation Recovery.
spacer

• View the DATABASE results for 'Saturation' (54).Open this link in a new window

Searchterm 'Spin Echo' was also found in the following services: 
spacer
News  (1)  Forum  (10)  
 
Segmented HASTE
 
A variation of the (Half Fourier Acquisition Single Shot Turbo Spin Echo) HASTE sequence, whereat half of the image information is acquired after the first excitation pulse, and the half after the second excitation pulse. The acquired data are then interleaved into the raw data matrix. A long time to repetition (TR) is selected to allow the spin system to recover between excitation pulses and the dead time is used for additional slices. The length of the echo train is cut in half. Also more than 2 segments are possible.
spacer

• View the DATABASE results for 'Segmented HASTE' (2).Open this link in a new window

MRI Resources 
Health - Education - NMR - Bioinformatics - RIS - Blood Flow Imaging
 
Sequential Line Imaging
 
MR imaging techniques in which the image is built up from successive lines through the object. In various schemes, the lines are isolated by oscillating magnetic field gradients or selective excitation, and then the NMR signals from the selected line are encoded for position by detecting the FID or spin echo in the presence of a magnetic field gradient along the line; the Fourier transformation of the detected signal then yields the distribution of emitted NMR signal along the line.
spacer

• View the DATABASE results for 'Sequential Line Imaging' (4).Open this link in a new window

MRI Resources 
Research Labs - Artifacts - MRI Technician and Technologist Jobs - Quality Advice - Breast Implant - Libraries
 
previous      91 - 95 (of 104)     next
Result Pages : [1 2 3 4 5 6]  [7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]