[Reply (1 of 2) to: 'How does TE affect DWI acquisition?' started by: 'danilo Scelfo' on Wed. 29 Sep.10]
Category:
Sequences and Imaging Parameters |
How does TE affect DWI acquisition? |
I never tried to shorten only the echo time while keeping all other parameters the same, but I would expect a change in SNR and artifact appearance. I would use parallel imaging methods to get a shortest possible echo time. Parallel imaging provides DWI with shorter echo trains associated with further reduction of the echo and repetition times, which reduces susceptibility effects and also image blurring. Additionally, the shorter echo time may lead to changes in ADC measurement.
|
|