Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 




 
MRI Forum
'bolus'
SEARCH FORUM FOR   
 
Result: Searchterm 'bolus' found in 5 messages
Result Pages: [1] 
More Results: Database  (24)  
Forum Overview
 bottom
Renate Bloemer

Mon. 13 Nov.06,
19:46

[Reply (1 of 3) to:
'MRA'
started by: 'Marco Costa'
on Sun. 22 Oct.06]


 
  Category: 
Protocols

 
MRA
A test bolus helps to get the delay at which time the CE-MRA sequence has to be started. The sequence needed is a one-slice dynamic fast gradient echo sequence (if possible with fat suppression or subtraction) and a temporal resolution of 0.5 to 1.5 sec. This thick slice is placed over the vessel of interest and the dynamic is started simultaneous with a small (1-2ml) bolus of Gad.
In the follow up you can see at which time the bolus is visible. If it is not clear to differentiate, a ROI measurement will help. The starting time depends on the k-space filling. Around the peak of the highest intensity, the contrast information should be read out (with "centric" at the beginning of the sequence).
Centric k-space filling order means that the central lines of the k-space are filled in the first seconds after starting the scan. Depending on the equipment, there is more than one non-centric order, for example linear, where the central lines are filled in the middle of the scan time.
Hope this helps.
 
 

Renate Bloemer
 View the whole thread
Marco Costa

Sun. 22 Oct.06,
10:27

[Start of:
'MRA'
2 Replies]


 
  Category: 
Protocols

 
MRA
Hello!

Is my first time in this forum! I read it a lot this forum, is very helpfull for my job!

I have a dificult in my mri machine.

First i would like to now how can i construct a bolus teste manualy! I don't have smart prep. What sequence i must use and what parameters i must choose.

Second, can you expline to me the diference between centric fill of K-space and non centric?

When is the best time to star the sequence after take the time injection.

My regards
Marco Costa
 View the whole threadReply to this thread
(login or register first)
Reader Mail

Sun. 19 Feb.06,
14:36

[Start of:
'CE-MRA Vasovist protocol'
1 Reply]


 
  Category: 
Protocols

 
CE-MRA Vasovist protocol
We will start soon with Vasovist (MS-325) for peripheral CE-MRA with moving table. Expecting higher resolution due to longer possible scan times - but discussing timing and needed post processing.

Is bolus timing needed - is fluoroscopic triggering possible - for the arterial phase?
Any experiences with this contrast agent, a good scan protocol or some good sources about Vasovist?
Anything appreciated!

Greetings
 View the whole threadReply to this thread
(login or register first)
Sue Johanson

Wed. 27 Apr.05,
03:36

[Reply (1 of 4) to:
'Contrast MRA COW.'
started by: 'Marilia White'
on Thu. 3 Mar.05]


 
  Category: 
Contrast Agents

 
Contrast MRA COW.
Hi,
seems crazy, you should have many undesirable veins! We run a 3D TOF MRA of the circle of willis without contrast with best results (1.5T and 3T).
Contrast enhanced 3D MRA of the carotids are done with machine injection (never with a butterfly!), automated bolus detection and a flow of 3ml/sec and 25 ml Gad. If no bolus detection program is available, look for the delay with a test bolus of 2ml. With a high resolution scan, both carotids and circle of willis are assessable. The use of a set like Smartset for the injection is helpful if no power injector is available.
 View the whole thread
Tomas Randell

Thu. 16 Dec.04,
05:03

[Reply (2 of 4) to:
'CE MRA of the Liver'
started by: 'bob mitchell'
on Sun. 5 Dec.04]


 
  Category: 
Applications and Examinations

 
CE MRA of the Liver
Like lily lily we use a muliphase MRA with 3 phases (artery, portal vein and late phase).
We use an amount of 30 ml Gadolinium with an injection rate of 5 ml/sec. With this high flow rate, you get a more concentrated bolus and therefore a higher contrast for the second phase. We use CENTRA k-space acquisition and reduce the scan time with SENSE (a parallel imaging technique like ASSET ), this will help for a better timing.
 View the whole thread

Result Pages : [1] 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]