Anna Lena Golay
Wed. 25 Jan.12, 20:59
[Reply (3 of 7) to: 'Imaging optic neuritis' started by: 'Karen Lesley' on Wed. 18 Jan.12]
Category:
General |
Imaging optic neuritis |
A 2-3 mm STIR sequence through the optic nerves may show the characteristic high-signal intensity of optic neuritis. Gadolinium enhanced T1-weighted fat-saturated (to suppress the bright signal of the orbital fat tissue) images of the orbits show the inflammation of the optic nerve. White matter lesions, which denote a higher risk of developing MS, are typical imaged with FLAIR and T2-weighted images (hyperintense lesions), or show enhancement of T1-weighted images postcontrast.
|
| View the whole thread | | |
Karen Lesley
Mon. 23 Jan.12, 11:02
[Reply (1 of 2) to: 'Gadolinium - blood brain barrier' started by: 'Reader Mail' on Mon. 23 Jan.12]
Category:
General |
Gadolinium - blood brain barrier |
Gadolinium cannot normally cross the bbb. It does when there is a breach though, and this shows up as a bright white spot on T1 images.
|
| View the whole thread | | |
Adana Osco Barras
Tue. 3 Jan.12, 21:09
[Reply (1 of 2) to: 'Hyperintense blood vessels in SE' started by: 'Peruvumba Jayakumar' on Thu. 29 Dec.11]
Category:
Sequences and Imaging Parameters |
Hyperintense blood vessels in SE |
Prolonged T1 relaxation times, decreased T1 tissue contrast, and increased susceptibility rnare typical effects at high field MRI.rnDue to the longer T1 times at 3T, the signal rnintensity inside the vessels is preserved even in smaller vessels. Try TR 700, TE 10.rnInversion recovery sequences are very well suited for optimal gray to white matter contrast, but enhancement after administration of gadolinium may not be visible.
|
| View the whole thread | | |
Peruvumba Jayakumar
Thu. 29 Dec.11, 06:38
[Start of: 'Hyperintense blood vessels in SE' 1 Reply]
Category:
Sequences and Imaging Parameters |
Hyperintense blood vessels in SE |
I am troubled by the blood vessels appearing bright on routine SE T1w images of the brain on 3T TIM system.
How can I avoid it?
Can I substitute T1 SE by T1 IR as a part of routine protocol?
|
| View the whole thread | Reply to this thread (login or register first) | |
kaisa bente
Thu. 27 Jan.11, 22:28
[Start of: 'Brain T2FS postcontrast' 1 Reply]
Category:
Applications and Examinations |
Brain T2FS postcontrast |
What is the advantage of T2 FS imaging of the brain if 3 planes of postcontrast T1's are obtained.
Thanks.
|
| View the whole thread | Reply to this thread (login or register first) |
| |