Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 




 
MRI Sequences
 
 
 
ball_redBalanced Sequence 
This family of sequences uses a balanced gradient waveform. This waveform will act on any stationary spin on resonance between 2 consecutive RF pulses and return it to the same phase it had before the gradients were applied. A balanced sequence starts out with a RF pulse of 90° or less and the spins in the steady state. Prior to the next TR in the slice encoding, the phase encoding and the frequency encoding direction, gradients are balanced so their net value is zero. Now the spins are prepared to accept the next RF pulse, and their corresponding signal can become part of the new transverse magnetization. If the balanced gradients maintain the longitudinal and transverse magnetization, the result is that both T1 and T2 contrast are represented in the image.
This pulse sequence produces images with increased signal from fluid (like T2 weighted sequences), along with retaining T1 weighted tissue contrast. Balanced sequences are particularly useful in cardiac MRI. Because this form of sequence is extremely dependent on field homogeneity, it is essential to run a shimming prior the acquisition.
Usually the gray and white matter contrast is poor, making this type of sequence unsuited for brain MRI. Modifications like ramping up and down the flip angles can increase signal to noise ratio and contrast of brain tissues (suggested under the name COSMIC - Coherent Oscillatory State acquisition for the Manipulation of Image Contrast).
These sequences include e.g. Balanced Fast Field Echo (bFFE), Balanced Turbo Field Echo (bTFE), Fast Imaging with Steady Precession (TrueFISP, sometimes short TRUFI), Completely Balanced Steady State (CBASS) and Balanced SARGE (BASG).

• View the DATABASE results for 'Balanced Sequence' (5).Open this link in a new window

 
Further Reading:
  News & More:
Generic Eddy Current Compensation for Rapid Magnetic Resonance Imaging(.pdf)
   by www.switt.ch    
Magnetic resonance imaging guided musculoskeletal interventions at 0.23T: Chapter 4. Materials and methods
2002
Balanced Fast Field Echo 
(bFFE) A FFE sequence using a balanced gradient waveform. A balanced sequence starts out with a RF pulse of 90° or less and the spins in the steady state. Before the next TR in the slice phase and frequency encoding, gradients are balanced so their net value is zero. Now the spins are prepared to accept the next RF pulse, and their corresponding signal can become part of the new transverse magnetization. Since the balanced gradients maintain the transverse and longitudinal magnetization, the result is, that both T1 and T2 contrast are represented in the image. This pulse sequence produces images with increased signal from fluid, along with retaining T1 weighted tissue contrast. Because this form of sequence is extremely dependent on field homogeneity, it is essential to run a shimming prior the acquisition. A fully balanced (refocused) sequence would yield higher signal, especially for tissues with long T2 relaxation times.

See Steady State Free Precession and Gradient Echo Sequence.

• View the DATABASE results for 'Balanced Fast Field Echo' (3).Open this link in a new window

 
Further Reading:
  News & More:
T1rho-prepared balanced gradient echo for rapid 3D T1rho MRI
Monday, 1 September 2008   by www.ncbi.nlm.nih.gov    
Utility of the FIESTA Pulse Sequence in Body Oncologic Imaging: Review
June 2009   by www.ajronline.org    
Balanced SARGE 

• View the DATABASE results for 'Balanced SARGE' (3).Open this link in a new window

Balanced Turbo Field Echo 
(BTFE) A gradient echo pulse sequence with a balanced gradient waveform and data acquisition after an initial preparation pulse for contrast enhancement.

See Steady State Free Precession (SSFP) and Balanced Sequence.

• View the DATABASE results for 'Balanced Turbo Field Echo' (3).Open this link in a new window

Fast Imaging with Steady Precession 
(TrueFISP) True fast imaging with steady state precession is a coherent technique that uses a fully balanced gradient waveform. The image contrast with TrueFISP is determined by T2*//T1 properties and mostly depending on TR. The speed and relative motion insensitivity of acquisition help to make the technique reliable, even in patients who have difficulty with holding their breath.
Recent advances in gradient hardware have led to a decreased minimum TR. This combined with improved field shimming capabilities and signal to noise ratio, has allowed TrueFISP imaging to become practical for whole-body applications. There's mostly T2* weighting. With the used ultrashort TR-times T1 weighting is almost impossible. One such application is cardiac cine MR with high myocardium-blood contrast. Spatial and temporal resolution can be substantially improved with this technique, but contrast on the basis of the ratio of T2* to T1 is not sufficiently high in soft tissues. By providing T1 contrast, TrueFISP could then document the enhancement effects of T1 shortening contrast agents. These properties are useful for the anatomical delineation of brain tumors and normal structures. With an increase in SNR ratio with minimum TR, TrueFISP could also depict the enhancement effect in myoma uteri. True FSIP is a technique that is well suited for cardiac MR imaging. The imaging time is shorter and the contrast between the blood and myocardium is higher than that of FLASH.

See Steady State Free Precession.

• View the DATABASE results for 'Fast Imaging with Steady Precession' (3).Open this link in a new window

 
Further Reading:
  Basics:
Accurate T1 Quantification Using a Breath-hold Inversion Recovery TrueFISP Sequence
2003   by rsna2003.rsna.org    
  Coherent Gradient Echo top
Human history becomes more and more a race between education and catastrophe.
- H. G. Wells
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]